The resolution profile during a fill will probably be slightly different for every magnet. For this particular magnet, it is advisable not to collect high resolution data for at least 30 minutes after a helium fill.
A blog for the NMR users at the University of Ottawa and all others interested in NMR spectroscopy.

The figure consists of simulations of X in an AA'X spin system as a function of JAA' with JAX set at 10 Hz and no coupling between A' and X. Clearly, the spectrum of X is affected by the coupling between A and A'. When JAA' = 0, a first order doublet is observed with a coupling constant of 10 Hz. As JAA' increases, complicated second order multiplets are observed. When JAA' = 50 Hz (or more) a "virtual triplet" with a coupling constant of 5 Hz is observed. This appears to be identical to a 1:2:1 triplet in a first order spectrum with a coupling constant of 1/2 JAX. It is however a second order spectrum and should not be misinterpreted as first order weak coupling. An example of this is illustrated in the figure below.
The figure shows the 13C NMR signals for the ipso and ortho aromatic carbons of 1,2-bis(diphenylphosphino)ethane (DPPE). These carbon atoms are coupled to the nearest phosphorus but not to the remote phosphorus. The two phosphorus atoms are strongly coupled to one another. The ortho carbons appear as a "virtual triplet" and the ipso carbons, a second order multiplet.
where Io is the intensity of B with no saturation of A, and I∞ , is the intensity of B when A is saturated for an infinite time. The saturation transfer effect is useful for situations where the exchange is slow on the NMR time scale but faster than (or of the same order as) T1B. An example using 31P NMR is illustrated in the figure below for a ruthenium phosphine complex which undergoes slow exchange between isomers with different modes of bonding. The 31P [1H] NMR spectrum is shown in the upper right-hand panel of the figure. In this case, the P atoms of isomer A are chemical shift equivalent and give a singlet while those for isomer B are chemical shift nonequivalent and give an AB pattern. The spectrum of isomer B is shown in the lower panel of the figure as a function of saturation time of isomer A. 


The second is by Victor Terskikh called nmr900 which posts news and information from Canada's showpiece high field solids NMR facility, The National Ultrahigh-Field NMR Facility for Solids.
Stay up-to-date and check them out!
It should also be noted that the overall intensity of the acetone signal decreases with respect to the chloroform signal as a function of the pulse duration due to the width of the excitation profile decreasing as a function of the pulse duration.

A sample of toluene and H2O/D2O was prepared. As these two liquids are immiscible, the sample is layered with the less dense toluene on top and the more dense water on the bottom. The bottom trace in the figure shows a conventional 1H NMR spectrum. Since the pulse used to collect the spectrum was a hard 90° pulse with a wide excitation profile, one can see both the toluene and the water in the spectrum. For the middle and upper traces, a field gradient of 48 G/cm was turned on while a 100 µs 90° Gaussian excitation pulse was applied. The only difference between the middle and upper traces is the offset frequency used in the excitation pulse. The middle trace represents the spectrum of a slice of the water in the bottom layer of the sample and the upper trace represents the spectrum of a slice of the toluene in the top layer of the sample.



A "sample" is prepared by pushing a NONMAGNETIC thermocouple through an NMR tube cap. The depth of the thermocouple is adjusted such that when the cap is put on the NMR tube, the tip of the thermocouple sits in the center of the rf coil. The NMR tube should contain a suitable liquid filled to the correct depth. The tube is placed in the spinner and set to the proper depth with a depth gauge. While holding onto the thermocouple, the sample is lowered into the magnet until it sits correctly in the NMR probe. The thermocouple is connected to a digital thermometer (some of these devices can use a second thermocouple in an ice water bath as a reference). The desired temperature is set on the variable temperature unit. When the temperature on both the variable temperature unit and digital thermometer have stabilized (~ 10 minutes), the values from each are recorded. This is repeated for temperatures over the desired temperature range and a calibration plot is constructed. Shimming is not an issue. Note that no NMR measurements are made and that the sample tube is not spinning.
The energy levels are no longer equally spaced. Furthermore, the position of the resonances resulting from the transitions depends on the orientation of the nucleus with respect to the magnetic field. Since all orientations are represented equally in a powder sample, each transition gives a powder spectrum. This is illustrated on the right-hand side of the figure. Note that the m = 1/2 and m = -1/2 energy levels are affected equally to first order and that the central (m = 1/2 - m = -1/2) transition is affected only by the second order perturbation. As a result the line width of the resonance from the central transition is much narrower than the other transitions. (In the figure, the central transition is clipped significantly.). Unlike the first order interaction, the second order interaction is field dependent. The width of the central transition depends inversely on the strength of the magnetic field.
The spectrum consists of a central transition, CT, and two satellite transitions, ST1 and ST2. Note that along with a lineshape due to the orientational dependence of the nucleus in the magnetic field, there is also an isotropic quadrupolar shift. The central and satallite transitions are not at the same frequency. This effect is completely separate from the chemical shift.

The cross-peaks highlighted in yellow in the top panel of the figure are expanded in the bottom panel. Black contours are positive and red contours are negative. The panel on the bottom left shows the cross-peaks for AX and BX. In the case of the AX cross-peak, the line drawn through the cross peak has a positive slope indicating that the passive couplings JAB and JBX are of opposite sign. The line drawn through the BX cross peak also has a positive slope indicating that the passive couplings JAB and JAX are of opposite sign. From this we can deduce that JAX and JBX are of the same sign. The panel on the bottom right shows the cross-peak for AB. In this case, the line drawn through the cross peak has a negative slope confirming that the passive couplings JAX and JBX are of the same sign. In conclusion, the geminal and vicinal coupling constants are of opposite sign.