It is not unusual to pay $50,000 - $120,000 for an NMR probe, depending on the configuration. Cryogenically cooled probes cost much more. Even the simplest NMR probes, which have very simple electronics, are very expensive. Here are some of the reasons why.
1. All of the components (including capacitors, coils, frame, shield, screws, adjustment rods, springs, supports etc...) must be nonmagnetic. These parts are often more expensive.
2. Since the market for NMR probes is quite small, specialty parts are manufactured in small batches making them more expensive.
3. Parts near the coil must be manufactured out of materials which will not give a background NMR signal or have huge magnetic susceptibility differences. These parts are often expensive.
4. Many parts must be machined to very strict tolerances. This is especially true for MAS probes but also true in high resolution liquids probes as well. The cost of very precise machining is very expensive.
5. Some materials which are desirable for NMR probes are very difficult to machine and require specialized expensive tools as well as highly paid and highly skilled machinists. An example of such a material is zirconia, used in many MAS probes.
6. A great deal of research and development on the part of the instrument company is invested in optimizing the design of the probe to improve rf handling, signal-to-noise ratio, lineshape, temperature handling etc... Instrument companies must compete with one another in this regard. These costs must be recovered.
7. Each probe must be individually NMR tested to ensure that it meets all of the specifications. This requires an NMR spectroscopist and an NMR spectrometer, neither of which come cheap.
Please note that I was not prompted by any sales person or instrument company to post this BLOG entry ...... and I do realize that instrument companies make a profit from the sale of NMR probes.
Showing posts with label probe. Show all posts
Showing posts with label probe. Show all posts
Friday, May 16, 2008
Thursday, March 27, 2008
What is the Difference Between a Broadband and an Inverse Broadband Probe?
I am sometimes asked, what is the difference between a broadband probe and an inverse broadband NMR probe? Both probes typically have a fixed proton channel, a tunable broadband channel and a fixed 2H lock channel. The difference lies in the coil positions with respect to the sample. Broadband probes are optimized to observe an X nucleus (typically all nuclei with frequencies between 15N and 31P). The broadband, X coil is positioned closest to the sample to maximize the filling factor and hence the X nucleus sensitivity. The 1H coil is outside of the X coil and can be used for either 1H observation or 1H decoupling. The filling factor for this coil is lower than that of the X coil. In an inverse broadband probe, the 1H coil is closest to the sample maximizing the 1H sensitivity. The X coil is outside of the 1H coil and can be used for either X nucleus observation or decoupling. The X nucleus in an inverse broadband probe has much less sensitivity than in a broadband probe. The choice of probe depends entirely on whether the user desires more 1H or more X sensitivity. For example, The 13C spectra of dilute samples should be run in a broadband probe whereas 1H - 13C or 1H - 15N HSQC spectra of dilute samples should be run in an inverse broadband probe.
Labels:
broadband probe,
inverse broadband probe,
probe
Monday, March 17, 2008
Probe Coil Geometry
The NMR sample sits inside a coil within the NMR probe. The coil is part of a tuned circuit which delivers radio frequency (RF) pulses to the sample and detects the RF signals from the sample. The RF fields delivered by and detected from the coil must be perpendicular to the static magnetic field, Bo. The most efficient way to do this is with a horizontal solenoid coil. Such coils are commonly used in NMR probes for solids. The probe must be removed from the magnet when samples are changed. A less efficient, but more convenient, Helmholtz coil is commonly used for liquids samples. Like horizontal solenoid coils, vertical Helmholtz coils deliver and detect RF fields perpendicular to Bo, but due to their geometry they allow samples to be inserted and ejected form the probe with a pneumatic system. The probe need not be removed from the magnet to change samples.