Thursday, February 8, 2018

How Much Presaturation Power is Needed?

Measuring 1H NMR spectra of samples in water (or mixtures of H2O and D2O) usually requires some form of solvent suppression of which selective presaturation is the most common method.  In this technique, a long low power selective pulse is applied before the high power excitation pulse (or composite pulse).  If too little power is used for presaturation, the water signal will not be sufficiently suppressed.  If too much power is used, one loses intensity of signals close to the water resonance and foregoes the quantitative nature of the NMR data.  The question then arises as to how much power is required for presaturation.  The figure below shows the 300 MHz 1H NMR spectra of a plant extract in H2O/D2O measured as a function of presaturation power.  The data were acquired on a Bruker AVANCE II console with the Bruker zgcppr pulse sequence using a two second presaturation pulse.  The power levels are reported in dB and Hz.
One can see that at higher presaturation powers, one loses intensity for peaks near the water signal.  In this case, where the nearest resonance of interest is 52 Hz from the water signal, one can obtain unattenuated signals with a presaturation field strength of 27 Hz.  This corresponds to 58 dB on this instrument.  

Thursday, January 4, 2018

Improved Solvent Suppression with Composite Pulses

A hard rf pulse delivers an rf field to the NMR sample inside of the coil. The rf field is not perfectly homogeneous nor does it end abruptly at the edge of the coil. When a long, selective, low power presaturation pulse is given to suppress a water signal, the water in the coil may be fully saturated whereas water outside of the coil will not be. Furthermore, unsaturated water outside of the coil may be outside of the carefully shimmed region of the magnet and give rise to a broadened residual signal when presaturation is used prior to a hard 90° excitation pulse. One long known* way to avoid this residual broad signal is to use a composite 90° pulse after presaturation.  One of the simplest such pulses is a (90°x - 90°y - 90°-x - 90°-y) composite pulse which is designed to excite sample inside of the coil but not outside of the coil. As a result, when it is used after a water presaturation pulse, it will not excite the broad signal from the unsaturated water outside of the coil and it will provide a spectrum with better presaturation performance. The figure below shows a small portion of the 1H spectrum of a plant extract in H2O/D2O.
The water signal is highlighted in pink. The left panel shows a conventional spectrum acquired with a 90° pulse. The center panel is a spectrum of the same sample where a two second low power presaturation pulse preceded the 90° hard pulse (Bruker pulse program = zgpr). One can see that most of the water is suppressed from the presaturation pulse however, a broad water signal remains from the water outside of the coil. The spectrum in the right-hand panel is the same as that in the center except that the 90° pulse was replaced with a composite 90° pulse (Bruker pulse program = zgcppr). Clearly, the broad signal from the unsaturated water outside of the coil is essentially gone providing a spectrum with much better water suppression.

*A. Bax.  J. Magn. Res. 65, 142 (1985).