Monday, December 21, 2009

What is this Holiday Treat?

Things are beginning to wind down at the University of Ottawa as the end of exams approaches and we all look forward to a few holidays. I would like to wish all readers a happy and safe holiday season.

I leave you with a little puzzle. The 13C MAS NMR spectrum of one of my favorite holiday treats is shown below. What is it? Leave a comment to this post with your guess. (hint: cross polarization was attempted but was quite inefficient).

Monday, December 14, 2009

Defining the Excitation Profile

The excitation profile of an rf pulse is determined by its Fourier transform. The Fourier transform of rectangular pulses of monochromatic radiation, typically used in NMR measurements, are (sin(x) /x) (or sinc(x)) functions. The sinc(x) function has a large central lobe with satellite lobes of alternating positive and negative sign. In order to obtain uniform excitation and therefore quantitative data, one must ensure that the excitation pulse is sufficiently short to allow the entire spectral width of interest to fit within a small region of the central sinc(x) lobe. The pulse must also have sufficient amplitude to produce a 90° rotation of the magnetization. The excitation profile of four pulses is shown in the figure below.The data were obtained by measuring a 300 MHz 1H NMR spectrum of HDO as a function of transmitter frequency. The power level for each of the pulses was set such that the pulses provided a 90° flip angle for an on-resonance signal. Each spectrum was phased independently. The first zero crossings of the sinc(x) function are at + 1/(PW) and -1/(PW) where PW is the duration of the pulse. It is therefore important that the spectral width of interest be less than ~1/(10PW) to ensure uniform excitation. One can see that a 10 µsec pulse provides essentially flat excitation across 40 kHz whereas 50, 100 and 200 µsec pulses do not.

Thursday, December 10, 2009

Variable Temperature NMR - Thermal Equilibrium

When doing variable temperature NMR, students often ask me how long they should wait for thermal equilibrium in their sample before collecting NMR data. The answer depends of course on the amount of gas flow around the sample and the temperature difference between the current and desired sample temperature. The position of the thermocouple in an NMR probe is typically right below the sample. It takes time between when the thermocouple reports the desired temperature and when the sample is at the desired temperature. During this time there is a large thermal gradient across the sample as well as convection currents which will affect the line width of NMR resonances. These effects are demonstrated in the figure below. For this measurement, the temperature of the probe was set to 50°C with an air flow of 800 L/hour. Once the thermocouple read 50°C, a sample of D2O was placed in the probe and 30 minutes was allowed to pass, after which the sample was presumed to be at thermal equilibrium. The lock was established and the magnet was then shimmed. The sample was removed and allowed to sit at room temperature for 30 minutes. It was then reintroduced to the probe at 50°C. 1H NMR spectra of the residual HDO were then collected at 30 second intervals for a period of 10 minutes. As soon as the room temperature sample is reintroduced to the warm probe, it begins to warm up. During the this time, the thermal gradients and convection currents are large and the line width is adversely affected. As the sample temperature approaches 50°C the thermal gradients are smaller and the line becomes narrower. After approximately 6 minutes the width of the line changes very little. The sample appears to be at thermal equilibrium after 10 minutes.

Wednesday, December 2, 2009

Purge Pulses and Spin Locking Pulses

Both spin locking pulses and purge pulses are very useful components of multipulse NMR experiments. Spin locking pulses are long pulses applied at the same phase as the transverse magnetization. While being applied, the magnetization is polarized along the static field of the spin locking pulse, B1, in the rotating frame. The magnetization is therefore locked to the axis of the applied pulse in much the same way that an equilibrium magnetization vector is locked to the static magnetic field, Bo. Purge pulses are long pulses applied at a phase 90° from the transverse magnetization. While the pulse is being applied, the transverse magnetization precesses about the static field of the pulse, B1, exactly like the way transverse magnetization precesses about Bo during a delay. If the pulse is long enough, the magnetization will dephase as a result of the B1 inhomogeneity of the rf pulse and be lost.

A long high power pulse can behave as both a spin locking pulse and a purge pulse as demonstrated in the vector diagram below. Imagine a spectrum consisting of two singlets. If the transmitter is set to the frequency of one of the singlets and a 90°x pulse is applied, both magnetization vectors are rotated to the -y axis. During a delay equal to one quarter of the reciprocal frequency difference between the singlets, the "on resonance" singlet will remain stationary while the "off resonance" singlet will rotate by 90° onto the x axis. If a long high power pulse is now applied along the y axis, it will behave as a spin locking pulse for the "on resonance" singlet and a purge pulse for the "off resonance" singlet. An example of this is shown in the figure below for a sample of methylene chloride and chloroform where the transmitter was set on the methylene chloride resonance. The top trace represents a simple one pulse measurement. The spectrum in the bottom trace was collected by applying a 90°x pulse followed by a delay equal to one quarter of the reciprocal frequency difference between the methylene chloride and chloroform. A 1 msec y pulse was then applied at the same power level as the 90° pulse followed by detection. One can see that the resonance of methylene chloride is unaffected compared to the one pulse measurement while that of the chloroform has been completely suppressed.

Monday, November 30, 2009

B1 Homogeneity

High resolution NMR spectroscopists spend a great deal of time shimming the magnet to ensure that the static magnetic field, Bo is homogeneous. This is because the transverse magnetization precesses about Bo. The Larmor equation implies that if there is a distribution in Bo across the volume of the sample, there will be a distribution of frequencies for each resonance line (i.e. the NMR resonances will be broad). The more homogeneous Bo across the sample volume, the sharper the NMR lines.

There is another field we must consider when doing NMR experiments - the magnetic field due to the RF pulse in the rotating frame of reference. In the rotating frame of reference, during the application of a pulse, an "on resonance" NMR line experiences an effective field, Beff equal to the magnetic field due to the pulse, B1. B1 is a static magnetic field in the rotating frame of reference. Due to the finite dimensions of the coil in the probe with respect to the sample, the B1 field will not be homogeneous across the entire volume of the sample. For example, a 90° pulse for the sample in the center of the coil will not be equal to a 90° pulse for the sample near the edges of the coil. While an x phased pulse is being applied to an equilibrium magnetization vector, the magnetization will precess about the x' axis in the rotating frame in the z-y' plane exactly like transverse magnetization precesses about the z axis in the x-y plane. While the magnetization precesses in the z-y' plane during the pulse, it is affected by the inhomogeneity in the B1 field. The inhomogeneity of the B1 field can be measured by doing a simple pulse calibration, applying longer and longer pulses well beyond that needed for a 360° pulse. After the pulse, the magnetization vectors precess again about Bo, and can be measured. The magnitude of the magnetization for a 90°, (90° + 360°), (90° + 720°) ..... etc. pulse will depend on the B1 homogeneity. An example of this is shown in the figure below. The figure shows a simple 1H pulse calibration for the decoupler coil of a 5 mm broadband NMR probe. The B1 homogeneity is expressed as the ratio of intensity for an 810° pulse compared to that from a 90° pulse. In this case the B1 homogeneity is 0.43. Much higher B1 homogeneity would be expected for an inverse detection probe.

Thursday, November 26, 2009

Why are MRI Scanners so Loud?

If you have ever had a magnetic resonance image (I prefer the term NMR image) taken in a hospital, you have undoubtedly heard very loud noises while the instrument collected the data. These noises are the result of the application of magnetic field gradients by way of the gradient coils inside the main magnetic field. The magnetic field gradients are applied in short pulses and are used to spatially encode the sample (person) such that the Larmor frequency of the protons in one region of the sample differs from that in other regions. It is this spatial frequency labelling which allows for the generation of the image. The time dependant magnetic field generated by the pulsed magnetic field gradients interacts with the main magnetic field with a force exactly like the one we are all familiar with when we move two magnets in close proximity to one another. Since the pulsed field gradient coils are held in a rigid form within the main magnet, they are unable to move any great distance however the sudden application of a gradient pulse generates a very strong force which physically slams the gradient coils inside the rigid form making a loud noise from the vibration. The larger the main magnetic field or the stronger the gradient pulses, the greater the force generated and the louder the noise.

These noises can sometimes be heard in high resolution NMR probes equipped with pulsed field gradients. Recently, a student approached me concerned that the NMR spectrometer was making a soft "tick" noise for every scan collected in his gradient COSY experiment. Immediately I thought perhaps the probe was arcing so I turned down the RF power. The noise persisted. Upon closer inspection, I found that the gradient strength was set unnecessarily high for the measurement. When the gradient strength was turned down, the "tick" noise ceased.

Friday, November 20, 2009

NMR and Food Chemistry - Popcorn

Both liquid state and solid state NMR have become very important tools in the food industry for both research and quality control. Since starch is a very important biopolymer and a major constituent in many foods, a great deal of work has been done on its chemical and physical characterization. Starches are 1 -4 linked polymers of glucose. Native starches are a mixture of a linear polymer (amylose) and a branched polymer (amylopectin). Furthermore, the starches are of two main types differing in their crystal structures and water content: the A type (cereal starches) and the B type (tuber starches). In addition to A or B starches, the starch granules in plants can have some amorphous starch as well. Corn starch is of the A type.

Popcorn is a snack enjoyed by millions around the world. As a child, I remember being fascinated at watching it pop and wondering what was going on. Much study has been devoted to the physics of popping corn. Essentially, every kernel of corn is a pressure vessel. When cooked, the moisture trapped within the starchy endosperm of the kernel is superheated. When the steam pressure inside the kernel becomes high enough the hull (pericarp) of the kernel explodes and the superheated water in the starch granules suddenly vaporizes, expands and rapidly cools making a solid foam out of the starchy endosperm. Each starch granule is a bubble in the solid foam.
Although much work has been done to understand starch and much work has been done to understand the popping of corn, I was unable to find any efforts directed to the chemical changes in corn starch before and after popping. This prompted me to collect a few spectra to address the issue. The first figure below shows the starch region of the 13C CPMAS spectra of unpopped corn (bottom trace), popped corn (middle trace) and cooked but unpopped corn (upper trace). The spectrum of unpopped corn is a mixture of A type starch (with a characteristic three line pattern in the C1 region) and amorphous starch (with a broad distribution of overlapping lines in the C1 region). The spectra of popped corn and cooked but unpopped corn are essentially identical and characteristic of amorphous starch. The data indicate that the heating and dehydration of the corn transform the crystalline A starch into amorphous starch. The observation is consistent with published studies on the hydration of starches. Aside from this change, there is no evidence for any other chemical transformation in the starch.

The second figure shows an expansion of the C1 region highlighting the conversion of crystalline A type starch into amorphous starch.

Thursday, November 12, 2009

Probe Arcing

Probe arcing can occur during the application of an rf pulse. It is the passage of a spark between a localized area of high voltage inside the probe to ground. In cases of severe probe arcing, one can hear a "snap" during the application of a pulse. If the probe is removed from the magnet and the cover is removed, one can see the arcing as a small "bolt of lightning" between the high voltage area and ground. When a probe arcs, the integrity of the pulse is affected and one will observe FIDs of irreproducible amplitude and phase. Needless to say, the quality of data collected on an arcing probe will be severely compromised. If a probe is permitted to arc for extended periods of time, some of the electronic components inside the probe can be permanently damaged. The amplifiers can also be damaged as arcing will cause mismatching and high levels of reflected power. Here are a few things you can do to help stop probe arcing.

1. Use lower pulse power levels.
2. If the location of the arcing can be found visually (i.e. you can see the spark), then the high voltage area and the path to ground can be wrapped with teflon tape.
3. Round off any sharp edges inside the probe as these are the areas of highest local voltage. In particular, any solder joints around the coil should be smooth.
4. Keep the coil and capacitors as far away from ground as possible.
5. Purge the inside of the probe body continuously with nitrogen gas.

Monday, September 28, 2009

Field Dependence of Chemical Shift Resolution in Solids

As undergraduates we learn that the chemical shift resolution in an NMR spectrum increases with increasing field strength. This is true because the number of Hz in a ppm increases with field and inequivalent resonances are more separated (in Hz) at higher field.

Will one always get better chemical shift resolution at higher field? You may be surprised to learn that the answer is - no. In certain cases for liquids, where dynamic processes are occuring at a rate comparable to the frequency difference between resonances (i.e. on the NMR time scale) one may even get less resolution at higher fields due to line width changes at higher field.

In the MAS or CPMAS spectra of solids one obtains liquid-like spectra of solid materials. In the cases where dipolar coupling is effectively removed by either or both MAS and high power decoupling, the width of the resonances often depends on a chemical shift distribution associated with a particular site. Just like chemical shielding anisotropy, this distribution increases (in Hz) with the magnetic field strength. As a result, the chemical shift resolution does not improve on going to higher field as the widths of the MAS lines increase (in Hz) as a function of field. Two examples of this are shown below. The first figure is the 29Si CPMAS spectrum of the clay kaolinite at 11.7 T (top trace) and 21.1 T (bottom trace). Both spectra were acquired with MAS rates exceeding the span of the chemical shift tensors. One can see that there is little if any improvement in the chemical shift resolution at higher field.
The second figure is the 19F MAS spectrum of the perfluorinated polymer, Nafion at 11.7 T (top trace) and 21.1 T (bottom trace). The MAS rates for each spectrum were chosen such that the dipolar coupling between the fluorines was effectively averaged by the MAS and that the spinning sidebands would be coincident (in ppm) in the spectra. Again, one can see that there is little if any improvement in the chemical shift resolution at higher field.

Thank you to Victor Terskikh and Eric Ye of the National Ultrahigh-Field NMR Facility for Solids for providing the figures for this post.

Tuesday, August 4, 2009

The 500 MHz 1H NMR Spectrum of Butane

n-Butane is a very simple molecule. Should it not then give a very simple 1H NMR spectrum? The figure below shows two calculated 500 MHz proton spectra for n-butane. Which spectrum most closely represents the true spectrum of n-butane?The answer is the very complicated spectrum B. The spectra were calculated with the following parameters:

Spectrum B
Spectrometer frequency = 500 MHz
δ1 = δ4 = 0.333 ppm
δ2 = δ3 = 1.271 ppm
3J12 = 3J34 = 7.11 Hz
4J13 = 4J24 = -0.07 Hz
3J23 = 6.77 Hz
LB = 0.5 Hz

Spectrum A
Spectrometer frequency = 500 MHz
δ1 = δ4 = 0.333 ppm
δ2 = δ3 = 1.271 ppm
3J12 = 3J34 = 7.11 Hz
4J13 = 4J24 = -0.07 Hz
3J23 = 0 Hz
LB = 0.5 Hz

The only difference between the simulations is that in spectrum B a coupling of 6.77 Hz was assumed between the two methylene groups whereas in spectrum A the same coupling was taken to be zero. The reason spectrum spectrum B is so complicated is that despite the fact that both the methyl groups and both the methylene groups are chemically equivalent, they are not magnetically equivalent. This is true for both spectrum A and spectrum B however, in spectrum A the second order effects are small based on the parameters used in the simulation.

Thank you to Adrian Dingle for inspiring me to create this post.

Friday, July 31, 2009

Distortions in QCPMG Spectra from Pulse Miscalibrations

A standard QCPMG NMR pulse sequence consists of a 90o pulse followed by a train of 180o pulses. Ideally, the resulting spikelet envelope should outline the static lineshape from a conventiaonal Hahn-echo experiment. If the first pulse deviates from 90o due to incorrect calibration, the QCPMG spikelet pattern does not change significantly, the only effect is somewhat lower overall intensity (first figure). At the same time the miscalibrated subsequent pulses lead to significantly distorted spikelet patterns (second figure). The 180o pulse misset by as little as 20o-30o, could produce considerable oscillations in the spikelet intensity across the envelope. This illustrates that QCPMG NMR experiments are much more sensitive to proper setup of the 180o pulses than the Hahn-echo experiment. The QCPMG spectra shown were calculated in SIMPSON for a central transition of a spin 3/2 nucleus resonating at 295 MHz (87Rb at 21.1 T), CQ=10 MHz, ηQ=0.7, CS anisotropy= -200 ppm, coincidental EFG and CSA tensors, ωRF/2π= 200 kHz.

Many thanks to Eric Ye of the National Ultra-high Field NMR Facility for Solids for contributing this post.

Tuesday, July 28, 2009

13C NMR with 1H and 31P Decoupling

NMR users are very familiar with the advantages of proton decoupling when observing 13C. The 13C NMR spectra of phosphorus containing compounds can be made simpler by applying 31P decoupling either on its own or in addition to proton decoupling. The figure below shows the 13C NMR spectrum of dimethyl methylphosphonate with all possible combinations of proton and 31P decoupling. The data collection required a triple resonance probe with the appropriate band pass filters.

Friday, July 10, 2009

31P - 13C HMQC

When most people think about HMQC or HSQC spectra they think about protons and 13C or protons and 15N. Although these are by far the most common spins to probe, the HMQC technique can be applied to other spin pairs as well. In earlier posts to this BLOG, this was demonstrated for protons and 11B and 31P and 109Ag. This post shows an application of the technique to 31P and 13C. Measurements made between spin pairs, where one of the spins is not a proton, require a triple resonance probe and the appropriate filters.
The figure below shows 31P detected 31P - 13C HMQC data for dimethyl methylphosphonate with 13C decoupling during the acquisition and 1H decoupling during the entire sequence. The spectra on the top and left-hand sides of the plots are separately run one dimensional 31P and 13C spectra, respectively, both with proton decoupling. The methyl carbon of dimethyl methylphosphonate has 1JCP = 142 Hz whereas the methoxy carbons have 2JCP = 6 Hz. Two HMQC spectra were run. The one on the left was optimized for 142 Hz and the one on the right was optimized for 6 Hz. Each spectrum shows a correlation according to its coupling constant. The data were collected on a 500 MHz instrument with the appropriate bandpass filters on each channel.

Thursday, June 4, 2009

90 Degree Pulse Determinations

In the routine procedure of R.F. pulse optimization, distorted nutation curves can sometimes be observed, as can be seen in the calculated graph below. This is usually due to the short recycle delay time (D1) not sufficient for complete relaxation. With increasingly shorter delay time, the maximum of the nutation curve shifts to lower flip angle, which makes the curve asymmetric. Under this circumstance, the 90 degree pulse determined from the maximum of the intensity will be inaccurate. On the other hand, if the recycle delay time is not too short, say 1 to 2 times of T1, the length of 180 degree pulse can still be determined with reasonable accuracy.
Many thanks to Eric Ye of the National Ultra-high Field NMR Facility for Solids for contributing this post.

Friday, May 15, 2009

The Selective 1D Gradient NOESY

Many students run very long 2D NMR experiments to find one particular piece of information. This is especially true for 2D NOESY experiments where perhaps only a single NOE correlation is sought. Depending on the amount of compound available, 2D NOESY measurements often take hours or tens of hours to acquire and can cost alot of money in instrument user fees. Many people do not realize that there are selective 1D analogs to the non selective 2D experiments. These 1D experiments rely on shaped pulses for selective excitation and take only a small fraction of the time required to run the comparable non selective 2D experiment. Previously I have posted entries in this BLOG on the very useful 1D selective gradient TOCSY experiment. The figure below demonstrates the use of the 1D selective gradient NOESY experiment for bis(phenylthio)methane compared to the standard proton NMR spectrum. In this case, the methylene protons were selectively irradiated and the NOE's were observed on the phenyl rings. The experiment took less than 5 minutes to acquire.

Wednesday, April 29, 2009

The Background from a Dirty NMR Probe

Have you ever wondered why the manager of your NMR equipment gets on your case about wiping down NMR tubes before putting them in the NMR probe? The figure below should answer your question. It shows the 1H NMR spectrum of a clean empty NMR tube inside an NMR probe before and after the NMR probe has been cleaned.
Over time, the "stuff" from your hands and residue on the outside of your NMR tubes builds up on the inside of the inserts inside the coil of the NMR probe. This "stuff" contains protons and results in a background signal in all subsequent NMR spectra. I have seen inserts of NMR probes so dirty that samples have "stuck" inside the probe. The offending gunk resembles grey-black bubble gum or perhaps dirty dried nasal effluent. Periodically NMR probes must be cleaned to remove this offensive residue. This can usually be accomplished by gently inserting and removing a cotton swab soaked in alcohol inside the coil insert. For the particular case in the figure above, the NMR probe could not be cleaned in this way and had to be disassembled by a service engineer and cleaned in an ultrasonic bath.

WIPE YOUR NMR TUBES BEFORE PUTTING THEM IN AN NMR PROBE !!

Thank you to Dr. Michael Lumsden who manages the NMR Facility at Dallousie University for suggesting this post, kindly sharing his tale and providing the figure.

Wednesday, March 18, 2009

Spin Echos for Uncoupled Spins

The spin echo is one of the most fundamental building blocks for NMR pulse sequences. Its main purpose is to refocus chemical shifts. The simplest spin echo is that for uncoupled spins where only the offset, Ω (i.e. the frequency difference between the carrier and the resonance) need be considered. The pulse sequence is represented in the upper portion of the figure with the vector and product operator representations below. A 90x pulse is first given to create magnetization along the -y axis of the rotating frame. During the first delay period, τ, the magnetization rotates in the x-y plane at a rate, Ω. The 180x pulse rotates the magnetization 180 degrees about the x axis. During the second delay period, the magnetization again rotates in the x-y plane at a rate, Ω in the same direction as during the first delay. At the end of the second delay, the magnetization is on the y axis and the collection of the FID is started. It is important to note that the echo will always have its maximum at 2τ after the 90 degree pulse regardless of its offset, Ω or the duration of τ. The value of τ however is limited by the T2.

Thursday, March 5, 2009

What is T1ρ and How is it Measured?

The time constant for the build up of magnetization along the direction of the main magnetic field, Bo, (the z axis) either after a pulse or upon initially exposing a sample to the magnetic field is called the T1 relaxation time or spin-lattice relaxation time. It is this relaxation time which determines the rate at which a pulse sequence can be repeated. The time constant for the decay of magnetization in the x-y plane of the rotating frame of reference after a pulse is called the T2 relaxation time, the spin-spin relaxation time or the transverse relaxation time. It is this relaxation time which determines the natural line width of a particular resonance. There is another relaxation time constant of interest to NMR spectroscopists - T1ρ. T1ρ is the time constant for the decay of magnetization along the radio frequency field, B1, of an applied spin locking pulse in the rotating frame of reference. It is analogous to T1 except it describes relaxation along the radio frequency field of the pulse (which is static in the rotating frame) rather than relaxation along Bo. T1ρ's are of interest in ROESY, TOCSY and cross polarization experiments. The T1ρ is measured by first applying a 90 degree pulse to an equilibrium magnetization vector. A spin locking pulse is then applied. The phase of this pulse is shifted 90 degrees with respect to the excitation pulse such that the field of the spin locking pulse is coincident with the spin vector in the rotating frame of reference. During the spin locking pulse, the large magnetization vector (which was initially polarized in Bo) decays to its equilibrium value in the much smaller field, B1, with time constant, T1ρ. The T1ρ is measured by analysing the intensity of the NMR signal in spectra collected as a function of the duration of the spin locking pulse. This is illustrated in the figure below.

Tuesday, January 27, 2009

Ottawa U Takes Delivery of a Bruker AVANCE III 400

The Bruker AVANCE III 400 NMR spectrometer for solids was delivered to Ottawa U last week and is currently being installed. It will be available for use in February. With the addition of this instrument, the Ottawa U campus boasts 7 NMR instruments (8, including the Bruker AVANCE II 900 on the NRC campus ). For some photos of the installation, follow this link.



Friday, January 23, 2009

Weak One-bond or Multiple Bond Correlations in 1H / 13C HMQC / HSQC Spectra

Many people are quite surprised to see either unusually weak one-bond correlations or weak multiple bond correlations in their 1H / 13C HMQC / HSQC spectra. These people must be reminded that there is nothing "magic" about these experiments - the responses are based solely on an assigned delay proportional a reciprocal coupling constant. The large scale success of the 1H / 13C HMQC / HSQC techniques can be attributed to the fact that most one-bond 1H - 13C coupling constants are very similar ( ~ 145 Hz). The pulse sequences are therefore run with a delay based on a 145 Hz coupling constant. When one-bond coupling constants are significantly different than 145 Hz then the correlation will be either very weak or absent in the spectrum. Also, if multiple bond couplings are unusually large then those multiple bond correlations may be present in the spectrum. The figure below is an example. In the 500 MHz HMQC spectrum of an alkyne (optimized for 145 Hz coupling), one can see an unusually small one-bond correlation between the terminal alkyne proton and its attached carbon. There is also a weak two-bond correlation between the terminal alkyne proton and the other alkyne carbon.

Wednesday, January 21, 2009

The Effect of Magic Angle Spinning and High Power 1H Decoupling on 13C Cross Polarization NMR Experiments

Cross polarization (CP), magic angle spinning (MAS) and high power 1H decoupling are all routine methods used in solid state NMR experiments. It is useful to see the effect of each of these techniques on a solid sample. The figure below shows 13C cross polarization NMR spectra of glycine at 4.7 Tesla collected with various combinations of magic angle spinning and high power 1H decoupling.The bottom spectrum was collected with neither MAS nor high power 1H decoupling. One can see two very broad overlapping lines due to the carbonyl and methylene carbons. The broadening is due to chemical shielding anisotropy and heteronuclear dipolar coupling between the 13C and both 1H and 14N. The second trace from the bottom was collected with high power 1H decoupling but no magic angle spinning. The spectrum contains two broad resonances with very informative line shapes. The high power 1H decoupling effectively removes the 13C - 1H heteronuclear dipolar interaction. The line shapes are determined from the chemical shielding anisotropy and 13C - 14N dipolar coupling interactions. The second trace from the top was collected with magic angle spinning at 4.5 kHz but no high power 1H decoupling. The spectrum apparently contains only one broad resonance with spinning sidebands. The magic angle spinning effectively removes the 13C chemical shielding anisotropy interaction. Although MAS does help average the 13C - 1H heteronuclear dipolar interaction, the averaging is not very effective at a speed of 4.5 kHz. Also, MAS only partially averages the 13C - 14N heteronuclear dipolar interaction. The resonances are therefore broadened out by residual heteronuclear dipolar coupling. The methylene resonance is broadened to such an extent that it does not show up in the spectrum at all. The top spectrum was collected with both MAS and high power 1H decoupling. One can see two very sharp resonances due to the carbonyl and methylene carbons. The 13C chemical shielding anisotropy and 13C - 1H heteronuclear dipolar coupling interactions are effectively removed by the MAS and high power 1H decoupling, respectively. Since MAS does not average J coupling and only partially averages dipolar coupling between a spin I = 1/2 and quadrupolar nucleus, the methylene carbon shows fine structure due to both J coupling and residual 13C - 14N dipolar coupling (see inset in yellow).

Friday, January 16, 2009

The BIRD Filter

Many modern NMR experiments exploit coupling interactions between protons and heteronuclei (eg. 13C). In such sequences the goal is to selectively observe the protons bound to 13C and suppress those bound to 12C. Since 13C is only 1 % naturally abundant, this means that 99% of the signal must be suppressed. One particularly simple scheme to accomplish this is the BIRD (BIlinear Rotation Decoupling) filter. The BIRD filter uses a heteroneuclear spin echo with delays equal to 1/(21JCH) to align the 1H(12C) and 1H(13C) spin vectors along the -y and y axes of the rotating frame of reference, respectively. The 180 degree phase difference between the 1H(12C) and 1H(13C) spin vectors allows a 90 degree pulse to align the these vectors on the -z and z axes, respectively. At this point the 1H(12C) spins are allowed to relax according to their T1 to the null point. A final 90 degree read pulse puts the 1H(13C) spins in the transverse plane for observation. The first of the two figures below demonstrates the use of the BIRD filter on the lineshape sample. The second figure shows a vector diagram explaining the sequence.

 

Wednesday, January 14, 2009

HMQC vs HSQC

Proton detected Heteronuclear Multiple Quantum Coherence (HMQC) and Heteronuclear Single Quantum Coherence (HSQC) are both NMR techniques used to correlate the chemical shift of the protons in a sample to a heteronucleus such as 13C or 15N via the J coupling interaction between the nuclei. Since both techniques essentially provide the same information - a correlation map between the coupled spins - students sometimes ask which of these two methods is better and which should they use routinely. The difference between the two techniques is that during the evolution time of an HMQC both proton and X magnetization (eg: X = 13C ) are allowed to evolve whereas in an HSQC only X magnetization is allowed to evolve. This means that an HMQC is affected by homonuclear proton J coupling during the evolution period while an HSQC is not affected as there is no proton magnetization during the evolution time. The homonuclear proton J coupling manifests itself as broadening in the X dimension. The top panel of the figure below shows the 7.05 T 1H /13C HMQC and HSQC spectra of menthol with an expansion of one of the resonances highlighted in yellow. One can see that the expanded cross peak of the HMQC is broader in the 13C dimension than that of the HSQC. The bottom panel of the figure shows the corresponding 13C projection spectra. One can see that the resolution is better in the projection of the HSQC compared to the HMQC. One might conclude that, due to the higher 13C resolution, it is always better to run an HSQC rather than an HMQC. This is definitely the case if all of the pulses are calibrated well, however since there are many more pulses in an HSQC compared to an HMQC, it is more susceptible to losses in signal-to-noise-ratio due to poor probe tuning or poor pulse calibration. My advice to students is that, if high 13C resolution is required, then make sure the pulses are calibrated well on a well tuned and matched probe and run an HSQC. If high 13C resolution is not critical then run an HMQC.

Thursday, January 8, 2009

Bloch-Siegert Shifts

Bloch Siegert shifts are frequency differences between NMR signals observed in the presence and absence an rf field applied during the acquisition time. The shifts arise because the applied rf field changes the effective magnetic field experienced by nearby resonances. The resonances are always displaced away from the frequency of the irradiating field. The shift is inversely related to the difference in frequency between the irradiation and the resonance and therefore is generally not observed when heteronuclear decoupling is applied. When homonuclear decoupling is employed these shifts can become significant and are typically used to calibrate the strength of the homonuclear decoupling field. One must be aware of these effects when reporting chemical shifts in homonuclear decoupling experiments. The figure below shows the effect of applying homonuclear decoupling fields of varying strength in the 300 MHz 1H NMR spectrum of dimethyl acetamide. One can see that the displacement of the resonances is away from the decoupling frequency and that the magnitude of the shift is inversely related to the frequency difference between the resonance and the irradiation frequency.

Monday, January 5, 2009

1H / 27Al TRAPDOR NMR of Kaolinite

TRAPDOR (TRAnsfer of Populations in DOuble Resonance) NMR (Grey and Vega, JACS 117, 8232 (1995)) is a solid state NMR technique where the effects of dioplar coupling between a quadrupolar nucleus and a spin I = 1/2 nucleus can be observed in the spectrum of the spin I = 1/2 nucleus. The technique relies on a rotor synchronized spin echo of the spin I = 1/2 nucleus with CW irradiation of the quadrupolar nucleus during the first echo delay period. The CW irradiation during a single rotor cycle behaves like an adiabatic frequency sweep as the quadrupolar frequencies vary over the course of the rotor cycle. The effects of dipolar coupling between the quadrupolar nucleus and the spin I = 1/2 nuclei, which are normally averaged by MAS, are reintroduced in the TRAPDOR measurement and the complete refocusing of the spin I = 1/2 NMR signal is prevented. The technique therefore can be used to determine whether or not a spin I = 1/2 nucleus is close in proximity to a quadrupolar nucleus. The figure below shows the 1H / 27Al TRAPDOR NMR spectrum of kaolinite at 11.7 Tesla. The top two traces are conventional rotor synchronized 1H Hahn echo spectra acquired with MAS rates of 12 kHz and 2.8 kHz, respectively. The bottom trace was acquired at the same spinning speed as the middle trace with CW irradiation of the 27Al during the first echo delay. One can see a very much reduced 1H echo indicating the presence of heteronuclear 1H - 27Al dipolar coupling.
This technique can be used to "find" quadrupolar neuclei which are "invisible" by direct detection due to their very large quadrupolar coupling constants.