The Information Content of an FID

The free induction decay (FID) is a function representing the decay of transverse magnetization as a function of time after the application of a pulse (or pulse sequence).  The NMR spectrum is obtained by Fourier transforming the FID.  All of the information in the NMR spectrum (line width, intensity, phase, line shape ....) is contained in the FID.  Knowledge of what part of the FID represents a particular property of an NMR spectrum allows a user to process the raw time domain data in a way that maximizes the quality and information content of the processed frequency domain spectrum.  The figure below shows the real component of the complex FID for a sample in which there is a single resonance.  The FID is labelled in terms of its information content.
The shape of the envelope of the overall decay defines the line shape of the NMR resonance.  Exponential decays yield Lorentzian line shapes.  The frequency in the FID represents the offset of the resonance from the carrier frequency or in other words, how far the resonance is from the center of the spectrum.  The higher the frequency, the further off-resonance the peak.  The FID for an on-resonance peak is a simple decaying exponential with no oscillation frequency present.  The first point of the FID contains the integrated intensity and phase information.  Since time and frequency are reciprocals of one another, the duration of the decay before it disappears into the noise determines the sharpness of the NMR resonances and ultimately the resolution in cases where there is more than one resonance in the spectrum.  The longer the decay - the narrower the lines.  The early portion of the FID (short time) defines the broad features (wide frequency distributions) whereas the the latter portion of the FID (long time) defines the sharp features (narrow frequency distributions).  Also, since the FID is a decaying function, the early portion has a higher signal-to-noise ratio than the latter portion.  With this knowledge its is very easy to understand how best to apply apodization functions to the raw data for the purpose line broadening and resolution enhancement.  Processing techniques such as zero filling, forward linear prediction, backward linear prediction and signal-to-noise optimization as well as spectral artifacts such as truncation errors, baseline roll and receiver saturation errors are easily understood with knowledge of the information content in the FID.   

3 comments:

Anonymous said...

Thanks for so detailed info about FID.

Juan Carlos Poveda Jaramillo said...

Hi everyone, i try to Tun 1D experimenta at different frequencies, 13C, 1H, 29Si, 31P, 27Al, 79Kr, but all FIDs shows spurious sígnals with overflow message during the acqyistiom and also when it finish. I try to change the RG parameter low down to 1, however the spikes or spurious signals papears again. Also, I change the wires breteen the probé and preamplifiers, also I use RF filters to observe the FID, also spuriousnsignas apears again.

Someone know about it?

Thanks in advance

JC Poveda

Glenn Facey said...

JC,
Are these magic angle spinning experiments? If so, I have seen spurious spikes from decoupled arcing or unstable spinning. Perhaps the rotor is touching the rf coil. Try turning off the decoupled and the magic angle spinning. If you are measuring spectra of liquids in a high resolution probe, there may be interference from the 2H lock channel, the vt heater or noise from gradients. Try disconnecting everything from the probe except the single cable used for your simple single pulse experiment. If this solves the problem, start connecting everything one at a time until you find the problem. If the problem persists, it may be noise from the console.

Good luck

Glenn