A "sample" is prepared by pushing a NONMAGNETIC thermocouple through an NMR tube cap. The depth of the thermocouple is adjusted such that when the cap is put on the NMR tube, the tip of the thermocouple sits in the center of the rf coil. The NMR tube should contain a suitable liquid filled to the correct depth. The tube is placed in the spinner and set to the proper depth with a depth gauge. While holding onto the thermocouple, the sample is lowered into the magnet until it sits correctly in the NMR probe. The thermocouple is connected to a digital thermometer (some of these devices can use a second thermocouple in an ice water bath as a reference). The desired temperature is set on the variable temperature unit. When the temperature on both the variable temperature unit and digital thermometer have stabilized (~ 10 minutes), the values from each are recorded. This is repeated for temperatures over the desired temperature range and a calibration plot is constructed. Shimming is not an issue. Note that no NMR measurements are made and that the sample tube is not spinning.Friday, June 25, 2010
Temperature Calibration - An Alternative Method
A "sample" is prepared by pushing a NONMAGNETIC thermocouple through an NMR tube cap. The depth of the thermocouple is adjusted such that when the cap is put on the NMR tube, the tip of the thermocouple sits in the center of the rf coil. The NMR tube should contain a suitable liquid filled to the correct depth. The tube is placed in the spinner and set to the proper depth with a depth gauge. While holding onto the thermocouple, the sample is lowered into the magnet until it sits correctly in the NMR probe. The thermocouple is connected to a digital thermometer (some of these devices can use a second thermocouple in an ice water bath as a reference). The desired temperature is set on the variable temperature unit. When the temperature on both the variable temperature unit and digital thermometer have stabilized (~ 10 minutes), the values from each are recorded. This is repeated for temperatures over the desired temperature range and a calibration plot is constructed. Shimming is not an issue. Note that no NMR measurements are made and that the sample tube is not spinning.Tuesday, June 22, 2010
Solid State NMR of Half Integer Quadrupolar Nuclei
The energy levels are no longer equally spaced. Furthermore, the position of the resonances resulting from the transitions depends on the orientation of the nucleus with respect to the magnetic field. Since all orientations are represented equally in a powder sample, each transition gives a powder spectrum. This is illustrated on the right-hand side of the figure. Note that the m = 1/2 and m = -1/2 energy levels are affected equally to first order and that the central (m = 1/2 - m = -1/2) transition is affected only by the second order perturbation. As a result the line width of the resonance from the central transition is much narrower than the other transitions. (In the figure, the central transition is clipped significantly.). Unlike the first order interaction, the second order interaction is field dependent. The width of the central transition depends inversely on the strength of the magnetic field.Magic angle spinning is a technique used by solid state NMR spectroscopists to obtain high resolution NMR spectra of solids. Magic angle spinning at infinite speed completely averages the first order quadrupolar interaction but only partially averages out the second order interaction. The energy level diagram for a spin I = 5/2 nucleus spinning infinately fast at the magic angle is shown in the figure below along with a simulated spectrum.
The spectrum consists of a central transition, CT, and two satellite transitions, ST1 and ST2. Note that along with a lineshape due to the orientational dependence of the nucleus in the magnetic field, there is also an isotropic quadrupolar shift. The central and satallite transitions are not at the same frequency. This effect is completely separate from the chemical shift.
In practice, we cannot spin at infinite speed, however, we can often spin at a rate fast with respect to the width of the central transition. If the quadrupolar coupling constant is small enough, the central transition will be observed and affected only by the second order interaction. The satellite transitions, affected by both the first and second order interaction, are observed as a manifold of spinning sidebands. The intensity of the centerbands for the satellite transitions is greatly attenuated as the overall intensity is spread among all of the sidebands. Often the centerbands for the satellite transitions are so small in comparison to that of the central transition that they are not observed. This is illustrated with simulations in the lower portion of the figure below. For comparison, the upper portion shows similar simulated spectra without magic angle spinning. The spectra highlighted in yellow are expansions of the central portion of the spectrum. These simulations are also supported by observations.
Thursday, June 17, 2010
Heat Dissipation in Bruker AVANCE Spectrometers
After years of service, two of our Bruker AVANCE spectrometers were running very warm. Using "UniTool" to check the temperature of the SGUs revealed that the board temperatures averaged 58° C! The DDS temperatures were all >55° C and not regulated. The air filters in the console doors were removed and cleaned and all eight fans of the AQS unit, housing the SGU's, were replaced on each spectrometer. After a day for the instruments to come to a thermal steady state, the SGU and DDS temperatures were again measured. The SGU board temperatures averaged 49°C. The DDS units in three of the five SGU's were now regulated properly at 55°C. The other two were still > 55°C and unregulated. In an attempt to lower the temperature further, the backs of the spectrometers were removed and again the instruments were allowed to reach a thermal steady state over a 24 hour period. The SGU and DDS temperatures were measured again. This time the SGU board temperatures averaged 38°C and all five DDS units were regulated properly at 55°C. The SGU temperatures in our AVANCE II and III spectrometers (with backs on) did not exceed 40°C and the DDS temperatures were regulated properly.
Although this may not be recommended by Bruker, I now run our AVANCE spectrometers with the back panels removed.
Friday, June 11, 2010
19F - 13C HMQC
Friday, June 4, 2010
E.COSY and the Relative Signs of Coupling Constants

The cross-peaks highlighted in yellow in the top panel of the figure are expanded in the bottom panel. Black contours are positive and red contours are negative. The panel on the bottom left shows the cross-peaks for AX and BX. In the case of the AX cross-peak, the line drawn through the cross peak has a positive slope indicating that the passive couplings JAB and JBX are of opposite sign. The line drawn through the BX cross peak also has a positive slope indicating that the passive couplings JAB and JAX are of opposite sign. From this we can deduce that JAX and JBX are of the same sign. The panel on the bottom right shows the cross-peak for AB. In this case, the line drawn through the cross peak has a negative slope confirming that the passive couplings JAX and JBX are of the same sign. In conclusion, the geminal and vicinal coupling constants are of opposite sign.1. C. Griesinger, O.W. Sorensen & R.R. Ernst, J. Magn. Reson. 75, ; 474 - 492 (1987).
2. The ecosygpph pulse program produces spectra similar to those described in reference 1 as "complimentary" E.COSY spectra. The slope of the lines through the cross peaks in "complementary" E.COSY spectra are of opposite sign to those obtained from the E.COSY spectra described in reference 1.