The signal-to-noise ratio in an NMR spectrum can be affected drastically the choice of the receiver gain setting, so care should be taken to set the receiver gain correctly for optimum results. At very low receiver gain settings, both the signal and the noise use only a fraction of the available digitization levels of the analog-to-digital concertor (ADC). As a result, the intensity of each point in the FID is represented with only a few possible values and the FID is "choppy". This is analogous to a black and white photograph being represented with a coarse gray scale of only a few shades of gray. Just like such a poorly represented photograph, the NMR spectrum contains a great deal of digital noise and therefore a low signal-to-noise ratio. As the receiver gain is increased, the FID is digitized with more available digitization levels. Since the thermal noise in the FID at low receiver gain settings is smaller than or comaparable the size of the digitization step of the ADC, the noise (unlike the signal) will not be amplified by increasing the receiver gain until it exceeds the size of the digitization step of the ADC after which it will be amplified in the same way as the signal. As a result, the signal increases more so than the noise as the receiver gain setting is increased therefore, the signal-to-noise ratio in the NMR spectrum increases steadily as the receiver gain setting is increased. As the receiver gain is increased beyond the point where the thermal noise exceeds the size of the digitization step, both the sginal and the noise can be digitized properly and the signal-to-noise ratio increases much less as a function of receiver gain setting increase. If the receiver gain is increased too much, the signal will exceed the limits of the ADC, the FID will be clipped at the beginning and the NMR spectrum will be severely distorted. The first figure below shows a series of spectra plotted as a function of the receiver gain setting. The spectra were scaled such that the signals were all of the same height. It is clear that the signal-to-noise ratio increases initially and then levels off. The data are plotted in the second figure.
Showing posts sorted by relevance for query receiver gain. Sort by date Show all posts
Showing posts sorted by relevance for query receiver gain. Sort by date Show all posts
Thursday, February 14, 2013
Monday, October 15, 2007
The Consequences of Setting the Receiver Gain Too High
The receiver gain of an NMR spectrometer is much like the volume control on a radio. When it is set too high there will be distortion in the NMR signal. The FID will be clipped near the beginning of the signal. The Fourier transform of this distorted signal is a distorted NMR spectrum. The figure below shows what this distortion looks like.
Many spectrometers will calculate the receiver gain automatically however you should be aware that this automatic calculation is not always perfect and that the receiver gain may have to be set manually. On a Varian spectrometer the receiver gain is the "gain" parameter. On a Bruker spectrometer the parameter is "rg". In both cases higher numbers mean a higher receiver gain.
Many spectrometers will calculate the receiver gain automatically however you should be aware that this automatic calculation is not always perfect and that the receiver gain may have to be set manually. On a Varian spectrometer the receiver gain is the "gain" parameter. On a Bruker spectrometer the parameter is "rg". In both cases higher numbers mean a higher receiver gain.
Friday, June 27, 2008
Artifacts Due To Setting the Receiver Gain Too High in 2D Homonuclear Experiments
Setting the receiver gain too high leads to very characteristic artifacts in 1D NMR spectra. If the receiver gain is set too high in 2D experiments, one can also expect artifacts in the 2D Fourier transformed data. In homonuclear experiments, setting the receiver gain too high will lead to parallel diagonal signals. This is illustrated in the COSY data in the figure below. In the left hand panel, the receiver was set correctly while in the right hand panel it was set too high. Dotted lines were drawn through the artifacts.
Friday, September 5, 2008
Backward Linear Prediction to Correct for Receiver Saturation
If the receiver gain is set too high the initial portion of the FID is clipped and the NMR spectrum is distorted. In such cases the spectrum should be run again with an appropriate receiver gain setting. Sometimes however, this may not be convenient as the sample may have decomposed. One way to improve the quality of the data is to take a close look at the FID, discard the initial clipped points, use backward linear prediction to calculate the discarded points and then do the Fourier transform. An example of the improvement you can expect is shown in the figure below.
Friday, October 17, 2008
Kinetic Experiments on Bruker Spectrometers
Students often have to monitor the progress of a chemical reaction as a function of time using NMR spectroscopy. I have written three simple programs for XWINNMR (which should work with little or no modification for TOPSPIN). Each program uses a different method to control the time allowed between collecting spectra. All are very simple and easily implemented. They should be added to the Bruker/XWINNMR/exp/stan/au/src directory. The first two programs, kinetic_ds and kinetic_t, are suitible for slow reactions where precise timing is not critical as they do not take into account the time required to initialize each acquisition. The third program, kinetic_2d avoids the problem by using a pseudo 2d approach and is suitible for faster reactions.
1. kinetic_ds
This program uses dummy scans to control the time allowed between spectra. (A dummy scan is a scan taken without turning on the receiver.) The more dummy scans, the longer the time between experiments. The user should set up the appropriate parameters and then run the program (by typing xau kinetic_ds). You will be asked for the total number of spectra to be collected, the number of scans to be collected for each spectrum and the number of dummy scans to be used in all but the first spectrum. The first spectrum will be collected in the current experiment and the others in subsequent experiments.
/* kinetic_ds */
/* written by Glenn Facey, August 24, 2005 */
/* This program will set up a kinetic run based on the use of dummy scans */
/* The user is asked for the number of spectra, the number of scans for */
/* each spectrum and the number of dummy scans for all but the first spectrum */
/* the first spectrum uses no dummy scans. */
GETCURDATA
GETINT("Enter total number of spectra",i1)
GETINT("Enter the number of scans for each spectrum",i2)
GETINT("Enter the number of dummy scans for all but the first spectrum", i3)
STOREPAR("ns",i2)
STOREPAR("ds",0)
Proc_err(0,"Kinetic Run in Progress");
RGA
ZG
TIMES(i1-1)
IEXPNO
SETCURDATA
STOREPAR("ds",i3)
STOREPAR("ns",i2)
ZG
END
QUITMSG("Data Collection Complete!")
2. kinetic_t
In this program, the user should set up the appropriate parameters and then run the program (by typing xau kinetic_t). You will be asked for the total number of spectra to be collected, the number of scans to be collected for each spectrum and the time in seconds allowed between the end of one acquisition and the beginning of the next acquisition. The first spectrum will collected in the current experiment and the others in subsequent experiments.
/* kinetic_t */
/* written by Glenn Facey, August 24, 2005 */
/* This program sets up and runs a kinetic experiment */
/* The user is asked to input the number of spectra, */
/* the number of scans for each spectrum and the time in */
/* seconds between the end of an acquisition and the */
/* beginning of the next. The program will measure the */
/* receiver gain and start the acquisitions. */
GETCURDATA
GETINT("Enter total number of spectra",i1)
GETINT("Enter the number of scans for each spectrum",i2)
GETINT("Enter the time interval (in seconds)", i3)
STOREPAR("ns",i2)
Proc_err(0,"Kinetic Run in Progress");
RGA
ZG
TIMES(i1-1)
IEXPNO
SETCURDATA
STOREPAR("ns",i2)
ssleep(i3);
ZG
END
QUITMSG("Data Collection Finished")
3. kinetic_2d
This program avoids initialization delays by collecting the data in a pseudo 2D format where each slice of the experiment is a spectrum. The program uses a pulse program called zg30kin.gf (see below) which should be put in the directory Bruker/XWINNMR/exp/stan/lists/pp (This pulse program program should be modified to suit the needs of the user). A variable delay list called kinetic must also be set up. This list contains the same number of lines as the number of spectra to be collected. Each line in the variable delay list defines the time interval (in seconds) to be allowed before each acquistion. The user must set up the appropriate parameters (including the number of scans to be collected for each spectrum) and then run the program (by typing xau kinetic_2d). You will be asked only for the total number of spectra to be collected. The program will set up a pseudo 2d acquisition. Data collection is started with the zg command. The data are processed with the xf2 command.
/* kinetic_2d */
/* written by Glenn Facey, August 24, 2005 */
/* This program sets up a pseudo 2D kinetic run */
/* using the pulse program zg30kin.gf with a Variable */
/* delay list called "kinetic". */
GETCURDATA
GETINT("How many spectra do you want to acquire?", i1)
FETCHPAR("SFO1",&d1)
FETCHPAR("DW",&f2)
FETCHPAR("SW",&d2)
FETCHPAR("SF",&d3)
XCMD("parmode 2D")
XCMD("pulprog zg30kin.gf")
XCMD("vdlist kinetic")
STOREPAR("SFO1",d1)
STOREPAR("DW",f2)
STOREPAR("SW",d2)
STOREPAR("SF",d3)
STOREPAR1("TD",i1)
STOREPAR1("SI",i1)
QUITMSG("Setup Complete!\n1. Define 'VD' List called 'kinetic'.\n2. Run the experiment with 'zg'.\n3. Process data with the 'xf2' command.")
Pulse program zg30kin.gf
;zg30kin.gf
;zg30 modified to run kinetic experiment in pseudo 2D mode
;using VD list
;avance-version (00/02/07)
;1D sequence
;using 30 degree flip angle
#include
"d11=30m"
1 vd
ze
2 d1
p1*0.33 ph1
go=2 ph31
d11 wr #0 if #0 ivd
lo to 1 times td1
exit
ph1=0 2 2 0 1 3 3 1
ph31=0 2 2 0 1 3 3 1
;pl1 : f1 channel - power level for pulse (default)
;p1 : f1 channel - 90 degree high power pulse
;d1 : relaxation delay; 1-5 * T1
;d11 : short delay for I/O
1. kinetic_ds
This program uses dummy scans to control the time allowed between spectra. (A dummy scan is a scan taken without turning on the receiver.) The more dummy scans, the longer the time between experiments. The user should set up the appropriate parameters and then run the program (by typing xau kinetic_ds). You will be asked for the total number of spectra to be collected, the number of scans to be collected for each spectrum and the number of dummy scans to be used in all but the first spectrum. The first spectrum will be collected in the current experiment and the others in subsequent experiments.
/* kinetic_ds */
/* written by Glenn Facey, August 24, 2005 */
/* This program will set up a kinetic run based on the use of dummy scans */
/* The user is asked for the number of spectra, the number of scans for */
/* each spectrum and the number of dummy scans for all but the first spectrum */
/* the first spectrum uses no dummy scans. */
GETCURDATA
GETINT("Enter total number of spectra",i1)
GETINT("Enter the number of scans for each spectrum",i2)
GETINT("Enter the number of dummy scans for all but the first spectrum", i3)
STOREPAR("ns",i2)
STOREPAR("ds",0)
Proc_err(0,"Kinetic Run in Progress");
RGA
ZG
TIMES(i1-1)
IEXPNO
SETCURDATA
STOREPAR("ds",i3)
STOREPAR("ns",i2)
ZG
END
QUITMSG("Data Collection Complete!")
2. kinetic_t
In this program, the user should set up the appropriate parameters and then run the program (by typing xau kinetic_t). You will be asked for the total number of spectra to be collected, the number of scans to be collected for each spectrum and the time in seconds allowed between the end of one acquisition and the beginning of the next acquisition. The first spectrum will collected in the current experiment and the others in subsequent experiments.
/* kinetic_t */
/* written by Glenn Facey, August 24, 2005 */
/* This program sets up and runs a kinetic experiment */
/* The user is asked to input the number of spectra, */
/* the number of scans for each spectrum and the time in */
/* seconds between the end of an acquisition and the */
/* beginning of the next. The program will measure the */
/* receiver gain and start the acquisitions. */
GETCURDATA
GETINT("Enter total number of spectra",i1)
GETINT("Enter the number of scans for each spectrum",i2)
GETINT("Enter the time interval (in seconds)", i3)
STOREPAR("ns",i2)
Proc_err(0,"Kinetic Run in Progress");
RGA
ZG
TIMES(i1-1)
IEXPNO
SETCURDATA
STOREPAR("ns",i2)
ssleep(i3);
ZG
END
QUITMSG("Data Collection Finished")
3. kinetic_2d
This program avoids initialization delays by collecting the data in a pseudo 2D format where each slice of the experiment is a spectrum. The program uses a pulse program called zg30kin.gf (see below) which should be put in the directory Bruker/XWINNMR/exp/stan/lists/pp (This pulse program program should be modified to suit the needs of the user). A variable delay list called kinetic must also be set up. This list contains the same number of lines as the number of spectra to be collected. Each line in the variable delay list defines the time interval (in seconds) to be allowed before each acquistion. The user must set up the appropriate parameters (including the number of scans to be collected for each spectrum) and then run the program (by typing xau kinetic_2d). You will be asked only for the total number of spectra to be collected. The program will set up a pseudo 2d acquisition. Data collection is started with the zg command. The data are processed with the xf2 command.
/* kinetic_2d */
/* written by Glenn Facey, August 24, 2005 */
/* This program sets up a pseudo 2D kinetic run */
/* using the pulse program zg30kin.gf with a Variable */
/* delay list called "kinetic". */
GETCURDATA
GETINT("How many spectra do you want to acquire?", i1)
FETCHPAR("SFO1",&d1)
FETCHPAR("DW",&f2)
FETCHPAR("SW",&d2)
FETCHPAR("SF",&d3)
XCMD("parmode 2D")
XCMD("pulprog zg30kin.gf")
XCMD("vdlist kinetic")
STOREPAR("SFO1",d1)
STOREPAR("DW",f2)
STOREPAR("SW",d2)
STOREPAR("SF",d3)
STOREPAR1("TD",i1)
STOREPAR1("SI",i1)
QUITMSG("Setup Complete!\n1. Define 'VD' List called 'kinetic'.\n2. Run the experiment with 'zg'.\n3. Process data with the 'xf2' command.")
Pulse program zg30kin.gf
;zg30kin.gf
;zg30 modified to run kinetic experiment in pseudo 2D mode
;using VD list
;avance-version (00/02/07)
;1D sequence
;using 30 degree flip angle
#include
"d11=30m"
1 vd
ze
2 d1
p1*0.33 ph1
go=2 ph31
d11 wr #0 if #0 ivd
lo to 1 times td1
exit
ph1=0 2 2 0 1 3 3 1
ph31=0 2 2 0 1 3 3 1
;pl1 : f1 channel - power level for pulse (default)
;p1 : f1 channel - 90 degree high power pulse
;d1 : relaxation delay; 1-5 * T1
;d11 : short delay for I/O
Friday, November 21, 2008
Before You Leave .....
This may seem to be a strange post .... a rant really...... but very important.
Over the years I have seen many students start a long acquisition (or series of acquisitions) on spectrometers and then immediately leave the lab. After a long lunch, an afternoon of playing billiards, a good night sleep or perhaps a weekend of skiing, they return to the lab and find no useful data waiting for them.
Why? ......
Well ....... perhaps the spectrometer was set up to run 4 rather than 20,000 scans, perhaps the receiver was saturated, perhaps the recycle delay was set to 1000 seconds rather than 2 seconds, perhaps the pulses were not set correctly, perhaps the spectral width was set too small, perhaps the probe was not tuned and matched, perhaps a delay was set to 10 seconds rather than 10 milliseconds. perhaps a typing error was made in the command to start the acquisition..... etc.
The NMR lab charges you for your time whether you get useful data or not, so it is important to be careful.
Before you leave the lab......
1. Double check the parameters in your experiment and for all queued experiments.
2. Query the spectrometer as to how long the experiment will take ("expt" (Bruker), "time" (Varian)) and ask yourself if the response makes sense.
3. Check the probe tuning and matching.
4. Make sure the receiver gain has been set correctly.
5. Look at the first few scans to make sure you have a signal.
Over the years I have seen many students start a long acquisition (or series of acquisitions) on spectrometers and then immediately leave the lab. After a long lunch, an afternoon of playing billiards, a good night sleep or perhaps a weekend of skiing, they return to the lab and find no useful data waiting for them.
Why? ......
Well ....... perhaps the spectrometer was set up to run 4 rather than 20,000 scans, perhaps the receiver was saturated, perhaps the recycle delay was set to 1000 seconds rather than 2 seconds, perhaps the pulses were not set correctly, perhaps the spectral width was set too small, perhaps the probe was not tuned and matched, perhaps a delay was set to 10 seconds rather than 10 milliseconds. perhaps a typing error was made in the command to start the acquisition..... etc.
The NMR lab charges you for your time whether you get useful data or not, so it is important to be careful.
Before you leave the lab......
1. Double check the parameters in your experiment and for all queued experiments.
2. Query the spectrometer as to how long the experiment will take ("expt" (Bruker), "time" (Varian)) and ask yourself if the response makes sense.
3. Check the probe tuning and matching.
4. Make sure the receiver gain has been set correctly.
5. Look at the first few scans to make sure you have a signal.
Monday, November 19, 2012
EZNMR at the University of Ottawa
The NMR Facility at the University of Ottawa is equipped with eight NMR spectrometers and has on the order of 100 hands-on users at the graduate and post-doctoral level. Like any university NMR facility, the users enter at varying knowledge and experience levels: from "What does NMR stand for?" to "How do I do a shearing transform for my 5QMAS data set?". Also, the attitude of the user's supervisors varies considerably. Some supervisors want their students to spend as little time in the NMR lab as possible by collecting all of their data in automation with a sample changer so they can maximize their time at the bench. Others want their students to learn how to collect the best possible data and fully understand the NMR measurements they make. There is no doubt that collecting NMR data under complete automation is incredibly time-efficient however, collecting data in this way teaches the student nothing about NMR measurements. On the other hand, learning to use NMR spectrometers manually, at the most fundamental level, to collect the best possible data, requires a great deal of knowledge (both general and instrument-specific) and although it is the most educationally rewarding, it certainly provides less overall sample throughput. In our facility, almost all students are first given 10 minutes of training, on how to collect NMR data under complete automation using our only fully automated instrument. Running an NMR spectrometer in this way requires absolutely no knowledge of NMR spectroscopy. Most users are also interested in using the other less automated instruments. These students are provided with as much training as they desire. The job of the NMR facility is to educate and satisfy the needs or each user. Doing so, requires finding a "happy medium" between complete automation and complete manual spectrometer operation and using that medium as a minimum training standard. For the last ten years or so, the "happy medium" used by the University of Ottawa to run four of it its Bruker NMR instruments is based on a customized button panel approach. We have written button panels specific to each instrument and call the option EZNMR. We have included EZNMR as an entry on the top TOPSPIN menu bar. Clicking the EZNMR option opens up a button panel like the one shown below, used on our AVANCE 500 spectrometer.
Students are of course encouraged to learn more about spectrometer operation than is available through the EZNMR buton panels.
Each button either issues a command, runs a macro or runs an "au program". Some of our students use exclusively this panel to collect their data. The advantage to using such a system is that the student must at least learn all of the steps involved with collecting the data. A typical EZNMR session involves simply following the button panel from top to bottom. If the probe contains a sample, it is ejected with the EJECT button. A new sample is lowered into the probe with the INSERT button. The deuterium lock is established by pressing the LOCK button which prompts the user for the solvent and then establishes the lock. The SHIM button first calls up a standard set of shims and then initiates a gradient shimming routine. After the magnet is shimmed, the user presses any one of the green buttons depending on which NMR measurement they intend to carry out. Pressing one of these buttons prompts the user to define a data set and then calls up a reasonable set of parameters into that data set. If desired, the user can change the number of scans or some of the parameters by pressing the SCANS or PARAMETERS buttons. The probe is then tuned using the TUNE button. The START button optimizes the receiver gain and begins collecting the data. Once started, the data can be processed at any time using the Proc 1D button or halted using the HALT button. We use a similar button panel for the commonly used 2D NMR experiments which can be called up from the 1D panel. It is shown here:
The advantages to using this system are:
- It is highly customizable for the hardware of each instrument as it is based on macros and "au programs".
- It can be added to as demands change.
- All NMR spectrometers using EZNMR look pretty much the same so instrument specific training is less of an issue.
- Students can begin running NMR experiments very quickly.
- Students are more likely to ask questions about each step and can learn at their own pace while maintaining high sample throughput.
- Its use is entirely optional.
- It is much more time-efficient than complete manual operation.
Wednesday, March 4, 2015
BLOG Index
University of Ottawa NMR Facility BLOG INDEX
March 2015
1D selective HOESY
1D TOCSY measurements
1H NMR with X nucleus
decoupling
2D correlation
spectroscopy (COSY)
2D NOESY measurements
2H NMR of liquids
ASCII file generation
Background signals
Background suppression
Baseline correction
Benchtop NMR
Chemical exchange agents
for spectral simplification
Chemical shift referencing
Chemical shift tensors
Chemical shifts
Concentration gradients
Coupling
Decoupler pulse
calibration
Decoupling
DEPT
DEPTQ
Diffusion and DOSY
Dynamic processes studied
by 1D NMR
Echoes
Floor vibrations
Food and drink
Free induction decay
HMBC experiments
measuring 19F – 13C coupling in 1H - 13C HMBC
HMQC, HSQC and edited HSQC
experiments
iPad / iPhone apps
Isotope effects
Linear prediction
Lock
Magnet
Magnet cryogen fills and
spectral quality
Magnetic resonance imaging
(MRI)
NMR of more than one isotope
NOE’s and decoupling
Paramagnetism
Probes
Pulsed field gradients
Quadrature
spikes
Receiver gain
Relaxation
Sample limitation
Scans
Shimming
Solid state 2H NMR
spectroscopy
Solvent effects
Solvent suppression
Spin simulations
t1 noise removal in 2D
spectra
Triple resonance
experiments
Variable temperature
measurements
Video tutorials
Subscribe to:
Posts (Atom)