Showing posts with label NOE. Show all posts
Showing posts with label NOE. Show all posts

Thursday, January 19, 2012

Sorting Out NOE's for Exchanging Rotamers

2D NOESY spectra contain cross peaks from both NOE interactions and peaks due to rotamers in slow exchange with one another on the NMR tine scale. For small molecules, the cross peaks resulting from slowly exchanging rotamers are of the same sign as the diagonal peaks. The NOE cross peaks, on the other hand, are of opposite sign compared to the diagonal peaks. When both types of correlations are present there may be more NOE correlations than expected. What follows is an example of this. The figure below shows a color coded chemical structure of a ruthenium complex with a color coded partial 1H NMR spectrum.It is obvious from the NMR spectrum that all of the signals from the color coded protons are doubled in the spectrum. One possible explanation for this is that there is a slow rotation about the ruthenium carbon bond indicated with the red curly arrow allowing for two possible nonequivalent rotamers. This is confirmed with the 2D 1H NOESY spectrum shown in the figure below with a 0.9 second mixing time. The spectrum clearly shows exchange peaks between corresponding pairs of 1H signals from each rotamer.The interesting thing to note from the NOESY spectrum is that each aromatic proton (pink) from a single rotamer shows NOE correlations to the methyl groups (blue and yellow) of both rotamers - not just those from a single rotamer. With this data, it is not possible to assign the subspectrum of a single rotamer. Presumably, the assignment could be made by collecting a 2D NOESY spectrum at low temperature where the rotation was completely frozen out or by collecting a 2D NOESY spectrum with a very short mixing time where the rotation would be limited. The problem with the former approach is that the solvent may freeze at a temperature too high to stop the bond rotation. The problem with the latter approach is that the NOE's would be much reduced due to the short mixing time and collecting a 2D data set with sufficient signal to noise ratio would take a great deal of time. Another approach is to collect selective 1D gradient NOESY spectra with selective excitation of the aromatic proton from each rotamer individually. These data are shown in the figure below for two different mixing times.Each spectrum is displayed in two parts. The left-hand panel is the aromatic region with the selectively excited resonance colored red and the right-hand panel is the aliphatic region showing the NOE correlations to the methyl groups. From the spectra collected with a 2 second mixing time, one can see that the selective excitation is no longer selective due to bond rotation during the long mixing time. One can see inverted peaks for the aromatic protons of both rotamers despite the fact that the 1H signal of only one rotamer was selectively excited. Furthermore, NOEs to the methyl signals from both rotamers are present. The spectra collected with only a 0.2 second mixing time, on the other hand, show very selective excitation. The time scale of the bond rotation is obviously longer than the 0.2 second mixing time. The spectra show only the NOEs between the selectively excited aromatic proton and the methyl groups from a single rotamer. The NOEs build up fast enough to be observed during the 0.2 second mixing time before rotation occurs. These data allow for the assignment of signals from each of the rotamers.

Thank you to Justin Lummiss of Dr. Fogg's research group for sharing this interesting system.

Tuesday, October 7, 2008

Positive NOE's and the Decision on Which Decoupling Mode to Use

Nuclei with positive gyromagnetic ratios, such as 13C, exhibit positive NOE's with nearby protons. When observing 13C directly with proton decoupling, the intensity of the resonances will be increased to an extent dependant on the magnitude of the NOE's and the amount of time over which they are allowed to build up. When quantitative results are not sought after, it is always best to collect the NMR data with decoupling during both the acquisition time and the relaxation delay so that the effect of the NOE's on the intensity of the lines is maximized. The figure below compares the 13C NMR spectra for formamide collected with inverse gated decoupling (left) and full decoupling (right).
The above spectra were collected at 11.7 tesla with 4 scans, using a recycle delay of 30 seconds and an acquisition time of 2.3 seconds. This effect of positive NOE's on the resonance intensity should be compared to that for negative NOE's.

Monday, October 6, 2008

Negative NOE's and the Decision on Which Decoupling Mode to Use

Nuclei with negative gyromagnetic ratios, such as 15N and 29Si, exhibit negative NOE's with nearby protons. When observing these nuclei directly with proton decoupling, the intensity of the resonances will be decreased to an extent dependant on the magnitude of the NOE's. Usually it is best to collect the NMR data for these nuclei with inverse gated decoupling so that the effect of the NOE's on the intensity of the lines is minimized. There are cases however where the magnitude of the NOE is so large that the intensity of the resonance collected with full decoupling becomes negative and even stronger than that observed with inverse gated decoupling. In such cases it is advantageous to leave the decoupler on 100 % of the time. As the figure below shows, this is indeed the situation with the 15N resonance of formamide where the magnitude of the signal is smaller when inverse gated decoupling decoupling is used.
The above spectra were collected at 11.7 tesla with 4 scans, using a recycle delay of 120 seconds and an acquisition time of 2.3 seconds.

Thursday, July 24, 2008

What are Those Positive Peaks in My NOESY Spectrum?

Students will sometimes ask me, "What are those positive off-diagonal peaks in my NOESY spectrum?". Since the NOESY pulse sequence is exactly the same as the EXSY sequence, a NOESY spectrum will show all possible correlations due to cross relaxation. These correlations include those from NOE's, chemical exchange and conformational (or rotational) exchange. For small molecules, if one phases the diagonal responses such that they are positive, the NOE's correlations will be negative and the exchange correlations will be positive. The answer to the question is therefore that the positive peaks are due to exchange. The figure below shows an example where three different types of cross peaks are visible. The molecule in the figure has two distinct rotational conformations in slow exchange with one another and gives a proton spectrum with every resonance doubled. Each resonance for one rotational conformation gives a positive (black) cross peak correlating it to the corresponding resonance of the other conformation. There is also a positive cross peak correlating the -NH- proton in the molecule to the residual water in the DMSO-d6 solvent. These protons exchange chemically with one another. Finally, the negative (red) correlations in the figure are due to NOE's.
Thank you to Jean-Gregoire Roveda of Dr. Beauchemin's group for giving me permission to use his spectrum as an example.