Showing posts with label relaxation time. Show all posts
Showing posts with label relaxation time. Show all posts

Monday, April 27, 2020

12C/13C Isotope Effects on 1H T1 Relaxation Times

What is the 1H T1 relaxation time of chloroform?  It seems like a simple enough question, but the answer is not so simple.  The relaxation rate for any proton is the sum of relaxation rates resulting from several different mechanisms (eg. homonuclear dipolar coupling, heteronuclear dipolar coupling, chemical shielding anisotropy, spin rotation etc...).  Each of these mechanisms of relaxation depends on dynamic effects and the extent to which those processes occur at the Larmor frequency.   Often, in proton-rich organic compounds, 1H T1 relaxation is dominated by the homonuclear dipolar coupling interaction.  For chloroform, with only a single proton, there can be no intra-molecular homonuclear 1H dipolar interaction and the 1H relaxation rate must depend on other mechanisms.  One of these mechanisms is the result of the heteronuclear dipolar coupling interaction.  For the 13C isotopologue of chloroform, one would expect a significant heteronuclear dipolar interaction between the directly bound 1H and 13C.   This interaction is absent in the 12C isotopologue and one would therefore expect the T1 relaxation time of 13CHCl3 to be much shorter than that of 12CHCl3.  This is illustrated in the figure below.
The 1H T1 relaxation times for both 12CHCl3 and 13CHCl3 were measured with the inversion recovery method for a degassed, dilute (1%) sample of chloroform in acetone-d6.  The inversion recovery delay was varied from from  1 sec. to 300 sec.  The recycle delay was 300 sec.  Relaxation is much more efficient for 13CHCl3 compared to 12CHCl3.  The T1 for 13CHCl3 is only 46% that of 12CHCl3, indicating the significance of the heteronuclear 1H - 13C dipolar coupling interaction as a relaxation mechanism.       

Thursday, March 26, 2015

Mixture Resolution in 13C CPMAS NMR

The recycle delay necessary to get the highest signal-to-noise ratio in a multi-scan 13C CPMAS NMR spectrum depends on the relaxation properties of the protons in the sample.  The protons in pure solid samples normally belong to a single homogeneous dipolar coupled network.  As a result, all of the protons in the coupled network have a common T1 relaxation time.  One would expect the same behavior for a mixture of compounds only if the components were mixed at the molecular level.  If the compounds are not mixed at the molecular level, the sample consists of domains of pure materials, each of which has a common proton T1.  If the proton T1's of the domains are significantly different, then one has a means of discriminating between the domains and hence the compounds of the mixture with 13C CPMAS NMR data.  The figure below illustrates this principle for a tablet of vitamin C ground into a powder.  The vitamin C tablet consists primarily of ascorbic acid for which the structure is shown in the figure.  The other major solid organic additives are hypromellose (hydroxypropyl methylcellulose), stearic acid (n-C17H35COOH), magnesium stearate and carnauba wax (a complex mixture of C26 to C30 acids, esters and alcohols).  When the tablet is ground up, the powder consists of ascorbic acid domains, stearic acid domains, magnesium stearate domains and carnauba wax domains.

13C CPMAS NMR spectra were acquired with a 30 second and a 2 second recycle delay and are shown in (a) and (b), respectively.  One can see that relative intensity of the components in the mixture depends on the recycle delay.  The proton T1 of the ascorbic acid is obviously longer than that of the other components of the mixture.  The spectra in (c) - (e) are linear combinations of (a) and (b).  The linear combination in spectrum (c) was created such that the ascorbic acid resonances were nulled.  The resulting spectrum is that of only the organic additives. The hypromellose resonances are in the 50 ppm to 110 ppm range.  The aliphatic resonances of the stearic acid, magnesium stearate and carnauba wax overlap in the 10 ppm to 50 ppm range and appear to have similar proton T1's.  The linear combination in spectrum (d) was created such that the aliphatic stearic and wax resonances were nulled.  The resulting spectrum is that of ascorbic acid and the inverted spectrum of the hypromellose.  The linear combination in spectrum (e) was created such that the hypromellose resonances were nulled. The resulting spectrum is that of the ascorbic acid with the stearic acid, magnesium stearate and carnauba wax additives.  This combination allows observation of the ascorbic acid with no overlapping resonances from the additives.

Wednesday, September 17, 2008

The Effect of Dissolved Oxygen on Relaxation Times

Proton T1 relaxation occurs, to a very large extent, due to inter-proton dipole-dipole interactions and their time dependence as a result of molecular motion. Potentially, there is also dipole-dipole interactions between protons and unpaired electrons which also contribute greatly to proton relaxation. In fact, for 13C NMR, paramagnetic materials are sometimes added to the sample to make relaxation more efficient. Dissolved molecular oxygen from air is a paramagnetic material that is often overlooked. Like other paramagnetic materials it contributes significantly to the relaxation rate of protons. It can be eliminated from NMR samples by simply bubbling nitrogen through the sample for a few minutes or freeze-pump-thawing the sample a few times. The following data show the effect of dissolved oxygen on the proton relaxation times of ethyl acetate in acetone-d6.

T1's for sample under air 
quartet = 8.7 seconds
singlet = 7.7 seconds
triplet = 6.9 seconds

T1's for sample under nitrogen
quartet = 17.7 seconds
singlet = 13.2 seconds
triplet = 12.5 seconds


One can see that the relaxation times are nearly 100% longer when oxygen is eliminated from the sample. The figure below shows the proton T1 inversion recovery measurement for the acetate methyl singlet of ethyl acetate for the same sample before and after nitrogen was bubbled through the solution.

Monday, July 28, 2008

T2 vs T2*

The T2 relaxation time is the exponential decay constant for transverse magnetization (i.e. magnetization in the xy plane). In principle, one should be able to measure the T2 relaxation time by applying a 90 degree pulse to create transverse magnetization and measuring the decay constant of the FID. In reality however, the decay rate of the FID is also affected by such things as magnetic field homogeneity, unresolved coupling, temperature gradients.....etc. Because of these effects, the decay constant of the FID is called T2* rather than T2. T2* is an instrumentally dependant parameter and it determines the line width of an NMR resonance. T2, on the other hand, is a physically meaningful parameter independent of field inhomogeneity, J coupling and other factors. It is measured with a 90-tau-180-tau-FID pulse sequence as a function of tau. T2 is always greater than or equal to T2*. The figure below compares T2 to T2* for the proton resonance of CHCl3 for the lineshape sample in a reasonably well shimmed 300 MHz magnet. The line shape specifications were 0.3 Hz (at 50%), 2.9 Hz (at 0.55 %) and 6.2 Hz (at 0.11%). Even in a well shimmed magnet, the T2 for CHCl3 is nearly 19 times longer than the T2*.

Wednesday, October 31, 2007

T1 Measurements and Estimation

T1 relaxation time measurements are usually done with a simple 180 -tau -90, inversion recovery pulse sequence (see figure). Tau is varied from a small value to a large value and a nonlinear regression is carried out to fit the best T1 value.These measurements can be very time consuming. One can get a reasonable estimate of the T1 much more quickly. Follow these simple steps:

1. Call up the pulse sequence "t1ir1d" (Bruker) or "s2pul" (Varian).
2. Set p1 and p2 (Bruker) or PW and P1 (Varian) to the 90 degree and 180 degree pulses , respectively.
Set the recycle delay, d1 (Bruker and Varian) to something you believe is much longer than the T1.
3. Set tau to a very small value (3 microseconds for example). Tau is d7 on a Bruker spectrometer or d2 on a Varian spectrometer.
4. Collect a spectrum and phase it such that all peaks are negative (one scan is often enough for protons). Store the phase correction.
5. Repeat step 3. increasing d7 (Bruker) or d2 (Varian) until the peak of interest is nulled. If the peak is negative, tau is too short. If it is positive, tau is too long.
6. The T1 of the peak of interest is the tau value for the null divided by the natural log of 2.