Friday, April 22, 2016

CEST - Chemical Exchange Saturation Transfer

Chemical Exchange Saturation Transfer (CEST) is a technique where one resonance, in slow exchange with a second resonance, is saturated with a selective low power pulse followed by a hard non-selective 90° pulse.  The intensity of the second resonance is then diminished due to the transfer of saturation from the first resonance as the result of  chemical exchange.  The figure below demonstrates this for a 25 mM solution of salicylic acid in H2O/D2O buffered at pH 7.
The left-hand panel of the figure is a stacked plot of extracted spectra collected in a pseudo 2D acquisition as a function of saturation frequency.  The saturation frequency was varied from an initial value of 20 ppm to a final value of -20 ppm in steps of 0.2 ppm.  The spectra are plotted such that only the water resonance is on scale.  One can see that the intensity of the water resonance dips when a saturation frequency of ~14 ppm is applied, corresponding to the resonance frequency of the –COOH and –OH protons of the salicylic acid (which appear to be in fast or intermediate exchange with one another).  The water resonance of course also dips to zero when a saturation frequency of ~4.7 ppm is used, corresponding to a simple presaturation of the water.  The right-hand panel of the figure is a plot of the integral of the water resonance as a function of saturation frequency, showing again a dip at ~14 ppm.

CEST is used in MRI to provide image contrast where a chemical exchange agent is introduced and images are collected with and without saturation of the exchange agent.  The difference provides an image enhanced by the presence of the chemical exchange agent.

Thank you to Dr. Mojmir Suchy of Prof. Adam Shuhendler’s group at the University of Ottawa for arousing my interest in the use of CEST for MRI and preparing the sample used in this post. 

Friday, January 7, 2011

"The MRI Photocopier"

A very interesting note* has recently been published in Concepts in Magnetic Resonance Part A by Joseph Hornak (generous provider of the two excellent free online text books, The Basics of NMR and The Basics of MRI ). In this work, it is demonstrated that an MRI scanner can be used to reproduce an image of printed text. The technique is based on the fact that the toner used to print text contains ferromagnetic particles and when placed in the uniform magnetic field of an MRI magnet, the magnetic field near the ferromagnetic toner is distorted. The "sample" is prepared by placing the printed page face up on a polycarbonate plate and covering it with a very thin sheet of polyethylene. The polyethylene is covered with water (doped with CuSO4 as a relaxation agent). The "sample" is placed in the magnet such that the plane of the page is perpendicular to the magnetic field. The text is "scanned" by taking a 2D proton MR image of a slice of the water near the surface of the polyethylene. The distortion in the magnetic field due to the ferromagnetic toner renders the text visible in the 2D magnetic resonance image. This is demonstrated in the figure below (from the reference) where a 2.5 mm thick slice above the polyethylene was imaged. The text was printed in #36 Arial font. The author speculates that this technique may be applied to the analysis of paintings.
* Joseph P. Hornak, "Magnetic Resonance Imaging of Printed Text", Concepts in Magnetic Resonance Part A. 36A, 347 (2010).

Thursday, January 6, 2011

Sample Slice Selection in NMR Spectroscopy

In MRI, field gradients are used routinely for slice selection while imaging a sample. Gradients are also finding applications in high resolution NMR spectroscopy employing fast data collection techniques using parallel data acquisition for multiple slices of the same sample. A good example of this can be found in Carlos' BLOG (and the references therein). The slice selection is accomplished by turning on a linear field gradient across the sample while applying an excitation pulse. While the gradient is on, the frequencies of each of the NMR resonances is spread out according to length of the sample and the strength of the field gradient across the sample. A particular individual slice of the sample can be selected by either modifying the strength of the gradient while using a semi-selective pulse of a given excitation frequency or by modifying the offset frequency of the semi-selective pulse for a constant gradient strength. Both the gradient strength and the excitation profile of the pulse determine the thickness of the slice selected. When the gradient is turned off and the receiver turned on, the FID representing only the spectrum of the selected slice of the sample is collected. An educational example is shown in the figure below. A sample of toluene and H2O/D2O was prepared. As these two liquids are immiscible, the sample is layered with the less dense toluene on top and the more dense water on the bottom. The bottom trace in the figure shows a conventional 1H NMR spectrum. Since the pulse used to collect the spectrum was a hard 90° pulse with a wide excitation profile, one can see both the toluene and the water in the spectrum. For the middle and upper traces, a field gradient of 48 G/cm was turned on while a 100 µs 90° Gaussian excitation pulse was applied. The only difference between the middle and upper traces is the offset frequency used in the excitation pulse. The middle trace represents the spectrum of a slice of the water in the bottom layer of the sample and the upper trace represents the spectrum of a slice of the toluene in the top layer of the sample.

Thursday, November 26, 2009

Why are MRI Scanners so Loud?

If you have ever had a magnetic resonance image (I prefer the term NMR image) taken in a hospital, you have undoubtedly heard very loud noises while the instrument collected the data. These noises are the result of the application of magnetic field gradients by way of the gradient coils inside the main magnetic field. The magnetic field gradients are applied in short pulses and are used to spatially encode the sample (person) such that the Larmor frequency of the protons in one region of the sample differs from that in other regions. It is this spatial frequency labelling which allows for the generation of the image. The time dependant magnetic field generated by the pulsed magnetic field gradients interacts with the main magnetic field with a force exactly like the one we are all familiar with when we move two magnets in close proximity to one another. Since the pulsed field gradient coils are held in a rigid form within the main magnet, they are unable to move any great distance however the sudden application of a gradient pulse generates a very strong force which physically slams the gradient coils inside the rigid form making a loud noise from the vibration. The larger the main magnetic field or the stronger the gradient pulses, the greater the force generated and the louder the noise.

These noises can sometimes be heard in high resolution NMR probes equipped with pulsed field gradients. Recently, a student approached me concerned that the NMR spectrometer was making a soft "tick" noise for every scan collected in his gradient COSY experiment. Immediately I thought perhaps the probe was arcing so I turned down the RF power. The noise persisted. Upon closer inspection, I found that the gradient strength was set unnecessarily high for the measurement. When the gradient strength was turned down, the "tick" noise ceased.

Thursday, March 20, 2008

Gradient Calibration - 1D MRI

When people think about magnetic resonance imaging (MRI), they often think about the huge whole body imagers in hospitals. NMR spectroscopists use one dimensional MRI in a specially prepared sample to calibrate the Z gradient strength of their spectrometers. The sample consists of a plastic disk with a precisely known thickness immersed in a column of water (see the figure below). 1D MRI is conceptually quite simple - a linear field gradient is applied during the collection of an FID. Since the magnetic field varies across the sample and the NMR frequency is proportional to the magnetic field strength, the resulting NMR spectrum represents a one dimensional image of the sample. The spectrum has a "notch" missing as a result of the plastic disk. The width of the "notch" is proportional to the applied gradient strength and the thickness of the plastic disk. If the thickness of the plastic disk is precisely known, then the strength of the applied gradient can be calculated. The spectrum is rounded at the edges as a result of the gradient strength falling off away from the center of the sample. A modified version of this experiment using an echo is used as a routine method of calibrating gradient strengths.

Thursday, November 29, 2007

Nyquist Fold-backs in Magnetic Resonance Images

Nyquist fold-backs are not just limited to 1D and 2D NMR data. They can also be observed in magnetic resonance images. The figures below are magnetic resonance images of seedless grapes acquired on our AVANCE 500. The Nyquist fold-backs are circled in blue.