The Fourier transform of a sinc function is a box function, so if a long low power sinc shaped rf pulse is used one obtains a narrow flat box shaped excitation profile. This is indeed the case as can be seen in the bottom trace of the figure below. The trace is composed of a series of 1H NMR spectra of H2O/D2O where a truncated, 200 msec, 10 cycle sinc shaped monochromatic rf pulse was applied with varying rf frequency offsets. One can see that the excitation profile is, to a good approximation, a narrow flat box function. The deviation from a flat box function is the result of the truncation of the sinc pulse. It should be noted that the phase of the resonances is not constant across the excitation profile. Sinc pulses are not frequently used for selective excitation because of the phase problem and the fact that very long, minimally truncated pulses must be used. A frequently used alternative to the sinc pulse for selective excitation is the Gaussian shaped pulse. The Fourier transform of a Gaussian is a Gaussian and one therefore will obtain a narrow Gaussian shaped excitation profile when a long low power Gaussian shaped pulse is used. This is shown in the top trace of the figure below. This trace is similar to the bottom trace except a 20 msec Gaussian shaped pulse (with truncation at 1 % of the total height) was used rather than a 200 msec sinc pulse. Although the excitation is not flat, the phase is constant across the excitation profile and the total duration of the pulse is 10 times shorter than the sinc pulse.
It should also be noted that the overall intensity of the acetone signal decreases with respect to the chloroform signal as a function of the pulse duration due to the width of the 

A sample of toluene and H2O/D2O was prepared. As these two liquids are immiscible, the sample is layered with the less dense toluene on top and the more dense water on the bottom. The bottom trace in the figure shows a conventional 1H NMR spectrum. Since the pulse used to collect the spectrum was a hard 90° pulse with a wide 




A "sample" is prepared by pushing a NONMAGNETIC thermocouple through an NMR tube cap. The depth of the thermocouple is adjusted such that when the cap is put on the NMR tube, the tip of the thermocouple sits in the center of the rf coil. The NMR tube should contain a suitable liquid filled to the correct depth. The tube is placed in the spinner and set to the proper depth with a depth gauge. While holding onto the thermocouple, the sample is lowered into the magnet until it sits correctly in the NMR probe. The thermocouple is connected to a digital thermometer (some of these devices can use a second thermocouple in an ice water bath as a reference). The desired temperature is set on the variable temperature unit. When the temperature on both the variable temperature unit and digital thermometer have stabilized (~ 10 minutes), the values from each are recorded. This is repeated for temperatures over the desired temperature range and a calibration plot is constructed. Shimming is not an issue. Note that no NMR measurements are made and that the sample tube is not spinning.
The energy levels are no longer equally spaced. Furthermore, the position of the resonances resulting from the transitions depends on the orientation of the nucleus with respect to the magnetic field. Since all orientations are represented equally in a powder sample, each transition gives a
The spectrum consists of a central transition, CT, and two satellite transitions, ST1 and ST2. Note that along with a lineshape due to the orientational dependence of the nucleus in the magnetic field, there is also an isotropic quadrupolar shift. The central and satallite transitions are not at the same frequency. This effect is completely separate from the chemical shift.

The cross-peaks highlighted in yellow in the top panel of the figure are expanded in the bottom panel. Black contours are positive and red contours are negative. The panel on the bottom left shows the cross-peaks for AX and BX. In the case of the AX cross-peak, the line drawn through the cross peak has a positive slope indicating that the passive couplings JAB and JBX are of opposite sign. The line drawn through the BX cross peak also has a positive slope indicating that the passive couplings JAB and JAX are of opposite sign. From this we can deduce that JAX and JBX are of the same sign. The panel on the bottom right shows the cross-peak for AB. In this case, the line drawn through the cross peak has a negative slope confirming that the passive couplings JAX and JBX are of the same sign. In conclusion, the geminal and vicinal coupling constants are of opposite sign.
This tool consists of a pole with a diameter slightly less than that of the spinner. On the end of the pole is a tapered prong with a diameter of 4 mm at the tip and 6 mm at the base. This tool is inserted in the top of the magnet and gently lowered until the prong gets wedged into the empty spinner. The spinner is removed with the tool. The entire operation takes less than 10 seconds and no reshimming is required.
A non-selective hard 90°x pulse is first given followed by a pair of identical pulsed field gradients sandwiching a soft selective 180° pulse about the y axis. The hard 90° pulse rotates all spin vectors onto the -y axis. During the first gradient pulse the spin vectors
The bottom trace shows a standard one-pulse measurement. The middle and top traces show results from a selective gradient spin echo measurement with the selective 180° pulse set for methylene chloride and acetone, respectively.
The bottom traces show the 11B [1H] NMR spectra for a dilute sample of NaBH4 and a "real" synthetic sample on the left and right, respectively. One can see an enormous background signal from both the NMR probe and the NMR tube. In the case of the "real" synthetic sample, the information from the spectrum is difficult or impossible to recover. The top traces show similar spectra acquired using the composite pulse. The only background signal remaining is that from the portion of the NMR tube inside the coil. This pulse sequence (without proton decoupling) is in the Bruker pulse program library called "zgbs". It is not exclusive to 11B.
The bottom trace shows a single scan 13C NMR spectrum of a concentrated solution of menthol in CDCl3 after the sample had sat in the magnet for a minute or so. The top trace shows a similar single scan acquisition preceded by 16 dummy scans. The relaxation delay was set to 1 second and the acquisition time for the FID was 1 second. 90° pulses were used in both spectra. One can see that some of the resonances in the top spectrum are attenuated in comparison to the bottom spectrum. The 13C signal from the CDCl3 (which has a long T1) is missing in the spectrum acquired with dummy scans. The dummy scans have selectively presaturated the solvent.

One can see that some of the NH's are greatly attenuated in the spectrum acquired with presaturation compared to the spectrum run using WATERGATE. The more attenuated the signal, the faster the chemical exchange for that particular NH with water.
Here, one can see the same exchange losses pointed out in the previous figure for the presaturation spectrum compared to the WATERGATE spectrum. In addition, one can see that the intensity of a very sharp peak marked in yellow (likely due to a CH proton from free histidine) is less intense in the WATERGATE spectrum compared to the presaturation spectrum. This loss is due to the fast diffusion of the small free amino acid compared to the very large protein. In conclusion, one must be aware of the differences between the two solvent suppression methods if quantitative results are being sought.
Here, the sample is subjected to continuous irradiation about the x axis. While being irradiated, the magnetization vector rotates in the z-y plane at a nutation frequency proportional to the pulse power. The magnetization on the -y axis is defined by a sine function. Fourier transformation of this magnetization gives an antiphase doublet centered at zero whose splitting Δ
Here each dwell period is divided up into a period for irradiation and a period for detection. The duty cycle for the irradiation is the fraction of time for which the sample is being irradiated. The magnetization is sampled when the power is off. As in the case for continuous irradiation, the magnetization vector still rotates in the z-y plane however, the rotation is slower as it is scaled according to the duty cycle. The duration of the 90° pulse is d/(2 Δ
The nutation spectrum was measured for HDO using a duty cycle, d = 0.10 and a power level of 

This sequence uses a simple two step phase cycle to subtract out the effect of the direct 13C Boltzmann magnetization. The first part of the sequence uses a (90°-y) pulse to return the CP enhanced magnetization to the z axis. The decay of the enhanced magnetization down to its Boltzmann value is followed using a (90°x) pulse with detection of signals on the -y axis. The second part of the sequence uses a (90°y) pulse to put the CP enhanced magnetization on the -z axis. The recovery of the enhanced inverted magnetization back to its equilibrium Boltzmann value is followed using a (90°x) pulse with detection of signals on the y axis. The addition of the first and second parts of the experiment by way of the phase cycle allows for a simple calculation of the 13C T1 with both the advantages of CP enhancement and the ability to collect more scans per unit time. An illustration of this method is shown in the figure below where the 13C T1's of glycine were measured. (The small peak in the spectrum is a spinning sideband of the carbonyl carbon)

where
The shielding scale increases to the right.